Factors Affecting Outcomes in Cochlear Implant Recipients Implanted With a Perimodiolar Electrode Array Located in Scala Tympani
Authors:
- 1 December 2016
- journal article
- basic science
- Published by Wolters Kluwer Health in Otology & Neurotology
- Vol. 37 (10) , 1662-1668
Abstract
Objective: To identify primary biographic and audiologic factors contributing to cochlear implant (CI) performance variability in quiet and noise by controlling electrode array type and electrode position within the cochlea. Background: Although CI outcomes have improved over time, considerable outcome variability still exists. Biographic, audiologic, and device-related factors have been shown to influence performance. Examining CI recipients with consistent array type and electrode position may allow focused investigation into outcome variability resulting from biographic and audiologic factors. Methods: Thirty-nine adults (40 ears) implanted for at least 6 months with a perimodiolar electrode array known (via computed tomography [CT] imaging) to be in scala tympani participated. Test materials, administered CI only, included monosyllabic words, sentences in quiet and noise, and spectral ripple discrimination. Results: In quiet, scores were high with mean word and sentence scores of 76 and 87%, respectively; however, sentence scores decreased by an average of 35 percentage points when noise was added. A principal components (PC) analysis of biographic and audiologic factors found three distinct factors, PC1 Age, PC2 Duration, and PC3 Pre-op Hearing. PC1 Age was the only factor that correlated, albeit modestly, with speech recognition in quiet and noise. Spectral ripple discrimination strongly correlated with speech measures. Conclusion: For these recipients with consistent electrode position, PC1 Age was related to speech recognition performance. Consistent electrode position may have contributed to high speech understanding in quiet. Inter-subject variability in noise may have been influenced by auditory/cognitive processing, known to decline with age, and mechanisms that underlie spectral resolution ability.Keywords
This publication has 54 references indexed in Scilit:
- The development of a modified spectral ripple testThe Journal of the Acoustical Society of America, 2013
- How Linguistic Closure and Verbal Working Memory Relate to Speech Recognition in Noise—A ReviewTrends in Amplification, 2013
- Pre-, Per- and Postoperative Factors Affecting Performance of Postlinguistically Deaf Adults Using Cochlear Implants: A New Conceptual Model over TimePLOS ONE, 2012
- Implications of Minimizing Trauma During Conventional Cochlear ImplantationOtology & Neurotology, 2011
- Verification of Computed Tomographic Estimates of Cochlear Implant Array PositionOtology & Neurotology, 2011
- Evidence for the Expansion of Adult Cochlear Implant CandidacyEar & Hearing, 2010
- Phonological processing in post-lingual deafness and cochlear implant outcomeNeuroImage, 2010
- Role of Electrode Placement as a Contributor to Variability in Cochlear Implant OutcomesOtology & Neurotology, 2008
- Spectral-Ripple Resolution Correlates with Speech Reception in Noise in Cochlear Implant UsersJournal of the Association for Research in Otolaryngology, 2007
- A Model for Cochlear Implant Electrode Insertion and Force Evaluation: Results with a New Electrode Design and Insertion TechniqueThe Laryngoscope, 2005