A Quality Control Check to Ensure Comparability of Stereophotogrammetric Data between Sessions and Systems
Authors:
- Kirsty ScottKSKirsty Scott
- Tecla BonciTBTecla Bonci
- Lisa AlcockLisa Alcock
- Ellen BuckleyEBEllen Buckley
- Clint HansenClint Hansen
- Eran GazitEGEran Gazit
- Lars SchwickertLSLars Schwickert
- Andrea CereattiAndrea Cereatti
- Claudia MazzàClaudia Mazzà
- on behalf of the Mobilise-D ConsortiumOCon behalf of the Mobilise-D Consortium
Open Access
Abstract
Optoelectronic stereophotogrammetric (SP) systems are widely used in human movement research for clinical diagnostics, interventional applications, and as a reference system for validating alternative technologies. Regardless of the application, SP systems exhibit different random and systematic errors depending on camera specifications, system setup and laboratory environment, which hinders comparing SP data between sessions and across different systems. While many methods have been proposed to quantify and report the errors of SP systems, they are rarely utilized due to their complexity and need for additional equipment. In response, an easy-to-use quality control (QC) check has been designed that can be completed immediately prior to a data collection. This QC check requires minimal training for the operator and no additional equipment. In addition, a custom graphical user interface ensures automatic processing of the errors in an easy-to-read format for immediate interpretation. On initial deployment in a multicentric study, the check (i) proved to be feasible to perform in a short timeframe with minimal burden to the operator, and (ii) quantified the level of random and systematic errors between sessions and systems, ensuring comparability of data in a variety of protocol setups, including repeated measures, longitudinal studies and multicentric studies.Keywords
Funding Information
- Innovative Medicines Initiative (820820)
- NIHR Sheffield Biomedical Research Centre (IS-BRC-1215-20017)
- Engineering and Physical Sciences Research Council (EP/K03877X/1, EP/S032940/1)
This publication has 22 references indexed in Scilit:
- Effects of the calibration procedure on the metrological performances of stereophotogrammetric systems for human movement analysisMeasurement, 2017
- Analysis of accuracy in optical motion capture – A protocol for laboratory setup evaluationJournal of Biomechanics, 2016
- Bilateral step length estimation using a single inertial measurement unit attached to the pelvisJournal of NeuroEngineering and Rehabilitation, 2012
- Systematic accuracy and precision analysis of video motion capturing systems—exemplified on the Vicon-460 systemJournal of Biomechanics, 2008
- Accuracy of an optical active-marker system to track the relative motion of rigid bodiesJournal of Biomechanics, 2007
- Human movement analysis using stereophotogrammetryGait & Posture, 2005
- A quasi-static method for determining the characteristics of a motion capture camera system in a “split-volume” configurationGait & Posture, 2002
- A spot check for estimating stereophotogrammetric errorsMedical & Biological Engineering & Computing, 2000
- The measurement of human motion: A comparison of commercially available systemsHuman Movement Science, 1999
- A procedure to validate three-dimensional motion assessment systemsJournal of Biomechanics, 1993