Accuracy, precision, and reproducibility of myocardial T1 mapping: A comparison of four T1 estimation algorithms for modified look‐locker inversion recovery (MOLLI)
- 4 December 2016
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 78 (5) , 1746-1756
Abstract
Purpose: To compare the accuracy and precision of four different T1 estimation algorithms for modified Look‐Locker inversion recovery (MOLLI).Methods: Four T1 estimation algorithms, including the original fit, inversion group (IG) fit, instantaneous signal loss simulation (InSiL), and Bloch equation simulation with slice profile correction (BLESSPC) were studied. T1 estimation accuracy, precision, reproducibility, and sensitivity to heart rate (HR), flip angle (FA), and acquisition scheme (AcS) variations were compared in simulation, phantom, and volunteer studies.Results: T1 estimation accuracy of IG (‐2.4% ± 3.9%) and original fit (‐3.2% ± 1.4%) were worse than BLESSPC (0.2% ± 1.5%) and InSiL (‐0.7% ± 2.1%). The original fit had the best precision for T1 from 409–1884 ms for the same FA (0.67% ± 0.16% versus 0.90% ± 0.23% using IG, 0.78% ± 0.11% using InSiL, 0.77% ± 0.12% using BLESSPC). BLESSPC generated the most consistent in vivo T1 values over different FAs and AcS, and the T1 estimation reproducibility was similar (P > 0.3) among the four methods when FA = 35°. When using FA = 50°, the reproducibility was significantly improved only when using BLESSPC (1.6% ± 0.9 versus 2.6% ± 1.9%, P < 0.05).Conclusion: BLESSPC has superior accuracy and is the least sensitive to FA, HR, and AcS variations. T1 estimation using BLESSPC and FA = 50° is superior to conventional MOLLI with FA = 35° in accuracy and precision. Further clinical studies in varying pathological conditions are warranted to confirm our findings. Magn Reson Med 78:1746–1756, 2017. © 2016 International Society for Magnetic Resonance in Medicine.Keywords
Funding Information
- the National Heart, Lung, and Blood Institute of the National Institutes of Health (R01HL127153)
This publication has 29 references indexed in Scilit:
- T 1 measurements in the human myocardium: The effects of magnetization transfer on the SASHA and MOLLI sequencesMagnetic Resonance in Medicine, 2013
- Modified look-locker inversion recovery T1 mapping indices: assessment of accuracy and reproducibility between magnetic resonance scannersJournal of Cardiovascular Magnetic Resonance, 2013
- Influence of Off-resonance in myocardial T1-mapping using SSFP based MOLLI methodJournal of Cardiovascular Magnetic Resonance, 2013
- T1 and extracellular volume mapping in the heart: estimation of error maps and the influence of noise on precisionJournal of Cardiovascular Magnetic Resonance, 2013
- Modified Look‐Locker T1 evaluation using Bloch simulations: Human and phantom validationMagnetic Resonance in Medicine, 2012
- Non-contrast T1-mapping detects acute myocardial edema with high diagnostic accuracy: a comparison to T2-weighted cardiovascular magnetic resonanceJournal of Cardiovascular Magnetic Resonance, 2012
- Assessment of Myocardial Fibrosis With Cardiovascular Magnetic ResonanceJACC, 2011
- Shortened Modified Look-Locker Inversion recovery (ShMOLLI) for clinical myocardial T1-mapping at 1.5 and 3 T within a 9 heartbeat breathholdJournal of Cardiovascular Magnetic Resonance, 2010
- Optimization and validation of a fully‐integrated pulse sequence for modified look‐locker inversion‐recovery (MOLLI) T1 mapping of the heartJournal of Magnetic Resonance Imaging, 2007
- Modified Look‐Locker inversion recovery (MOLLI) for high‐resolution T1 mapping of the heartMagnetic Resonance in Medicine, 2004