Repeatability of Aberrometry-Based Automated Subjective Refraction in Healthy and Keratoconus Subjects
Open Access
- 29 October 2020
- journal article
- research article
- Published by Wiley in Journal of Ophthalmology
- Vol. 2020, 1-7
Abstract
Purpose. To compare the intersession repeatability of the Eye Refract, a new instrument to perform aberrometry-based automated subjective refraction, on healthy and keratoconus subjects. Materials and Methods. A cross-sectional and randomized study was performed. A total of 64 participants were evaluated in the study, selecting one eye per participant randomly. The sample was divided into two different groups: 33 healthy subjects (38.85 ± 13.21 years) and 31 with keratoconus (37.29 ± 11.37 years). Three refractions per participant with the Eye Refract were performed on three different days, without cycloplegia. The repeatability analysis of refractive variables (M, J0, and J45), binocular corrected distance visual acuity (BCDVA), and spent time in refraction was performed in terms of repeatability (Sr), its 95% confidence interval (r), and intraclass correlation coefficient (ICC). Results. There were no statistically significant differences ( ) between sessions in both groups for all refractive variables (M, J0, and J45) and BCDVA. Spent time in refraction was reduced as the sessions went by ( ). The Eye Refract was more repeatable for refractive errors assessment in healthy subjects (M : Sr = 0.27 D; J0 : Sr = 0.09 D; J45 : Sr = 0.06 D) compared to those with keratoconus (M : Sr = 0.65 D; J0 : Sr = 0.29 D; J45 : Sr = 0.24 D), while it was similar for BCDVA. Conclusions. The Eye Refract offered better repeatability to assess refractive errors in healthy subjects compared to those with keratoconus. Despite measurements being also consistent in keratoconus subjects, they should be treated with caution in clinical practice.Keywords
This publication has 31 references indexed in Scilit:
- Clinical Accuracy of the Nidek ARK‐1 AutorefractorOptometry and Vision Science, 2019
- What is the appropriate gold standard test for refractive error?Ophthalmic and Physiological Optics, 2017
- Accuracy of WASCA Aberrometer Refraction Compared to Manifest Refraction in Chinese Adult MyopesJournal of Refractive Surgery, 2009
- Influence of fogging lenses and cycloplegia on open‐field automatic refractionOphthalmic and Physiological Optics, 2008
- A Comparison of Autorefraction and Subjective Refraction With and Without Cycloplegia in Primary School ChildrenAmerican Journal of Ophthalmology, 2006
- Accuracy, Repeatability, and Clinical Application of Spherocylindrical Automated Refraction Using Time-Based Wavefront Aberrometry MeasurementsOphthalmology, 2006
- Accuracy of the WASCA Aberrometer Refraction Compared to Manifest Refraction in MyopiaJournal of Refractive Surgery, 2006
- A Comparison of Autorefractor PerformanceOptometry and Vision Science, 2004
- Clinical Evaluation of the Topcon BV‐1000 Automated Subjective Refraction SystemOptometry and Vision Science, 2004
- Comparison of Spherical Equivalent Refraction and Astigmatism Measured with Three Different Models of AutorefractorsOptometry and Vision Science, 2004