An optimized b-value distribution for triexponential intravoxel incoherent motion (IVIM) in the liver
Open Access
- 1 April 2021
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 85 (4) , 2095-2108
Abstract
Purpose To find an optimized b-value distribution for reproducible triexponential intravoxel incoherent motion (IVIM) exams in the liver. Methods A numeric optimization of b-value distributions was performed using the triexponential IVIM equation and 27 different IVIM parameter sets. Starting with an initially optimized distribution of 6 b-values, the number of b-values was increased stepwise. Each new b-value was chosen from a set of 64 predefined b-values based on the computed summed relative mean error of the fitted triexponential IVIM parameters. This process was repeated for up to 100 b-values. In simulations and in vivo measurements, optimized b-value distributions were compared to 4 representative distributions found in literature. Results The first 16 optimized b-values were 0, 0.3, 0.3, 70, 200, 800, 70, 1, 3.5, 5, 70, 1.2, 6, 45, 1.5, and 60 in units of s/mm(2). Low b-values were much more frequent than high b-values. The optimized b-value distribution resulted in a higher fit stability compared to distributions used in literature in both, simulation and in vivo measurements. Using more than 6 b-values, ideally 16 or more, increased the fit stability considerably. Conclusion Using optimized b-values, the fit uncertainty in triexponential IVIM can be largely reduced. Ideally, 16 or more b-values should be acquired.Keywords
This publication has 37 references indexed in Scilit:
- The effect of low b‐values on the intravoxel incoherent motion derived pseudodiffusion parameter in liverMagnetic Resonance in Medicine, 2014
- Comparison of breathhold, navigator-triggered, and free-breathing diffusion-weighted MRI for focal hepatic lesionsJournal of Magnetic Resonance Imaging, 2012
- Liver diffusivity in healthy volunteers and patients with chronic liver disease: Comparison of breathhold and free‐breathing techniquesJournal of Magnetic Resonance Imaging, 2011
- Toward an optimal distribution of b values for intravoxel incoherent motion imagingMagnetic Resonance Imaging, 2011
- Optimization of b‐value sampling for diffusion‐weighted imaging of the kidneyMagnetic Resonance in Medicine, 2011
- An in vivo verification of the intravoxel incoherent motion effect in diffusion‐weighted imaging of the abdomenMagnetic Resonance in Medicine, 2010
- Influence of cardiac motion on diffusion-weighted magnetic resonance imaging of the liverMagnetic Resonance Materials in Physics, Biology and Medicine, 2009
- Respiratory-Triggered Versus Breath-Hold Diffusion-Weighted MRI of Liver Lesions: Comparison of Image Quality and Apparent Diffusion Coefficient ValuesAmerican Journal of Roentgenology, 2009
- Measurement of signal‐to‐noise ratios in MR images: Influence of multichannel coils, parallel imaging, and reconstruction filtersJournal of Magnetic Resonance Imaging, 2007
- Practical approaches to the evaluation of signal‐to‐noise ratio performance with parallel imaging: Application with cardiac imaging and a 32‐channel cardiac coilMagnetic Resonance in Medicine, 2005