Heat generation caused by ablation of dental hard tissues with an ultrashort pulse laser (USPL) system
Authors:
- 12 May 2013
- journal article
- research article
- Published by Springer Nature in Lasers in Medical Science
- Vol. 30 (2) , 475-481
Abstract
Heat generation during the removal of dental hard tissues may lead to a temperature increase and cause painful sensations or damage dental tissues. The aim of this study was to assess heat generation in dental hard tissues following laser ablation using an ultrashort pulse laser (USPL) system. A total of 85 specimens of dental hard tissues were used, comprising 45 specimens of human dentine evaluating a thickness of 1, 2, and 3 mm (15 samples each) and 40 specimens of human enamel with a thickness of 1 and 2 mm (20 samples each). Ablation was performed with an Nd:YVO4 laser at 1,064 nm, a pulse duration of 9 ps, and a repetition rate of 500 kHz with an average output power of 6 W. Specimens were irradiated for 0.8 s. Employing a scanner system, rectangular cavities of 1-mm edge length were generated. A temperature sensor was placed at the back of the specimens, recording the temperature during the ablation process. All measurements were made employing a heat-conductive paste without any additional cooling or spray. Heat generation during laser ablation depended on the dental hard tissue (enamel or dentine) and the thickness of the respective tissue (p < 0.05). Highest temperature increase could be observed in the 1-mm thickness group for enamel. Evaluating the 1-mm group for dentine, a significantly lower temperature increase could be measured (p < 0.05) with lowest values in the 3-mm group (p < 0.05). A time delay for temperature increase during the ablation process depending on the material thickness was observed for both hard tissues (p < 0.05). Employing the USPL system to remove dental hard tissues, heat generation has to be considered. Especially during laser ablation next to pulpal tissues, painful sensations and potential thermal injury of pulp tissue might occur.Keywords
This publication has 30 references indexed in Scilit:
- Precise ablation of dental hard tissues with ultra-short pulsed lasers. Preliminary exploratory investigation on adequate laser parametersLasers in Medical Science, 2012
- Heat generation caused by ablation of restorative materials with an ultrashort pulse laser (USPL) systemLasers in Medical Science, 2011
- High-speed scanning ablation of dental hard tissues with a λ = 9.3 μm CO2 laser: adhesion, mechanical strength, heat accumulation, and peripheral thermal damageJournal of Biomedical Optics, 2011
- Influence of the degree of dentine mineralization on pulp chamber temperature increase during resin-based composite (RBC) light-activationJournal of Dentistry, 2010
- Thermal Safety of Er:YAG and Er,Cr:YSGG Lasers in Hard Tissue RemovalPhotomedicine and Laser Surgery, 2009
- Effect of simulated pulpal microcirculation on intrachamber temperature changes following application of various curing units on tooth surfaceJournal of Dentistry, 2009
- Temperature Response in the Pulpal Chamber of Primary Human Teeth Exposed to Nd:YAG Laser Using a Picosecond Pulsed RegimePhotomedicine and Laser Surgery, 2006
- Thermographic investigation of contemporary resin-containing dental materialsJournal of Dentistry, 2005
- Heat transfer and phase change during picosecond laser ablation of nickelInternational Journal of Heat and Mass Transfer, 2002
- Investigation and spectral analysis of the plasma-induced ablation mechanism of dental hydroxyapatiteApplied Physics B, 1994