Feasibility and accuracy of low-angle, non-orthogonal biplane fluoroscopic 2D–3D registration for dynamic analysis of the thumb carpometacarpal joint
Authors:
- Ryoya ShiodeRSRyoya Shiode
- Satoshi MiyamuraSMSatoshi Miyamura
- Tasuku MiyakeTMTasuku Miyake
- Hiroki KondoHKHiroki Kondo
- Toru IwahashiTIToru Iwahashi
- Hiroyuki TanakaHTHiroyuki Tanaka
- Yoshito OtakeYOYoshito Otake
- Yoshinobu SatoYSYoshinobu Sato
- Tsuyoshi MuraseTMTsuyoshi Murase
- Seiji OkadaSOSeiji Okada
- Kunihiro OkaKOKunihiro Oka
- 1 September 2026
- journal article
- Published by IOP Publishing in Medical Engineering & Physics
- Vol. 147 (9) , 095020
Abstract
Background. Accurate three-dimensional dynamic assessment of the thumb carpometacarpal (CMC) joint is important for understanding the pathomechanics of CMC osteoarthritis and evaluating treatment outcomes. Intensity-based biplane 2D–3D registration is a promising approach for dynamic bone motion analysis; however, conventional orthogonal biplane configurations are often difficult to implement for small-joint imaging in clinical settings. Methods. We validated the accuracy of intensity-based biplane 2D–3D registration for thumb CMC joint kinematic analysis using a phantom model under a low-angle, non-orthogonal biplane fluoroscopic configuration (53°). Registration-derived bone poses were compared with radiostereometric analysis-derived ground truth across 252 biplane fluoroscopic image pairs during palmar and radial abduction tasks. Translational and rotational errors were calculated for the thumb proximal phalanx, first metacarpal, trapezium, and second metacarpal. Results. The method achieved submillimeter mean overall component-wise translational errors for all evaluated bones (range: 0.11–0.31 mm). Mean overall component-wise rotational errors were below 3° for all bones (range: 0.44–2.64°). Larger axis-specific rotational root mean square errors were observed for the trapezium and thumb proximal phalanx, reflecting bone- and axis-dependent error patterns. Conclusion. Intensity-based biplane 2D–3D registration can provide sufficient accuracy for dynamic analysis of the thumb CMC joint using a low-angle, non-orthogonal biplane configuration. This approach may serve as a clinically feasible platform for assessing small-joint kinematics in settings where conventional orthogonal biplane systems are difficult to implement, supporting future in vivo studies of dynamic thumb CMC joint motion.Keywords
Funding Information
- Japan Society for the Promotion of Science (JSPS), Japan
- KAKENHI (JP 23K08612)
This publication has 22 references indexed in Scilit:
- Thumb motion is not the same as first carpometacarpal joint motionJournal of Biomechanics, 2025
- The passive biomechanics of the thumb carpometacarpal joint: An in vitro studyJournal of Biomechanics, 2024
- Disease severity versus pain severity: Range of motion differences during single‐ and multiplanar tasks in women with carpometacarpal osteoarthritisJournal of Orthopaedic Research, 2024
- In-vivo kinematics of the trapeziometacarpal joint in dynamic pinch motion using four-dimensional computed tomography imagingSkeletal Radiology, 2023
- Analysis of forearm rotational motion using biplane fluoroscopic intensity-based 2D–3D matchingJournal of Biomechanics, 2019
- Osteoarthritis of the carpometacarpal joint of the thumb: a new MR imaging technique for the standardized detection of relevant ligamental lesionsSkeletal Radiology, 2014
- Functional Task Kinematics of the Thumb Carpometacarpal JointClinical Orthopaedics and Related Research, 2014
- In Vivo Kinematics of the Thumb Carpometacarpal Joint During Three Isometric Functional TasksClinical Orthopaedics and Related Research, 2014
- In Vivo Pilot Study Evaluating the Thumb Carpometacarpal Joint During Circumduction2013
- Current Concepts of the Anatomy of the Thumb Trapeziometacarpal JointThe Journal of Hand Surgery, 2011