Effects of clinical use of NiTi reciprocating instruments on cyclic and torsional resistance, and on roughness
Authors:
- Edson Jorge Lima MoreiraEMEdson Jorge Lima Moreira
- Henrique dos Santos AntunesHAHenrique dos Santos Antunes
- Victor Talarico Leal VieiraVVVictor Talarico Leal Vieira
- Daniele Moreira CavalcanteDCDaniele Moreira Cavalcante
- Henrique Eduardo OliveiraHOHenrique Eduardo Oliveira
- Diogo da Silva OliveiraDODiogo da Silva Oliveira
- Gustavo De-DeusGDGustavo De-Deus
- Emmanuel João Nogueira Leal da SilvaESEmmanuel João Nogueira Leal da Silva
Open Access
- 1 January 2021
- journal article
- research article
- Published by FapUNIFESP (SciELO) in Brazilian Oral Research
- Vol. 35, e021
Abstract
Abstrac The aim of the present study was to evaluate the cyclic fatigue, torsional resistance and surface roughness of Reciproc R25 instruments in four different situations, namely as new instruments and as instruments tested after clinical preparation of one, two or three maxillary molars with four root canals. The total time required to perform each root canal preparation was recorded. Cyclic fatigue resistance was determined by the time to fracture using a customized testing device (n = 10 per group). The torsional test evaluated the torque and angle of rotation to failure according to ISO 3630-1 (n = 10 per group). The roughness of the working parts of new and used instruments was evaluated with a profilometer (n = 5 per group). Statistical analysis was performed using one-way ANOVA and Tukey’s test. The level of significance was set at 5%. No fractures or deformations were observed after clinical use. Higher preparation time was needed during the third use of the instruments for all root canals (p < 0.05). There were no significant differences among the groups in regard to either cyclic fatigue or torsional resistance (p > 0.05). Regarding the roughness measurements, groove depth was higher on new and one- versus two- or three-maxillary-molar-prepared instruments (p < 0.05). It can be concluded that the clinical use of Reciproc instruments increased preparation time and decreased surface roughness. However, clinical use did not affect the cyclic fatigue or torsional resistance of the Reciproc instruments.Keywords
This publication has 29 references indexed in Scilit:
- Influence of Kinematics on the Cyclic Fatigue Resistance of Replicalike and Original Brand Rotary InstrumentsJournal of Endodontics, 2020
- Nickel–titanium instruments in endodontics: a concise review of the state of the artBrazilian Oral Research, 2018
- New thermomechanically treated NiTi alloys – a reviewInternational Endodontic Journal, 2018
- Cyclic fatigue using severely curved canals and torsional resistance of thermally treated reciprocating instrumentsClinical Oral Investigations, 2018
- Accumulated Hard Tissue Debris Produced during Reciprocating and Rotary Nickel-Titanium Canal PreparationJournal of Endodontics, 2015
- Is the reciprocating movement per se able to improve the cyclic fatigue resistance of instruments?International Endodontic Journal, 2013
- Influence of Different Angles of Reciprocation on the Cyclic Fatigue of Nickel-Titanium Endodontic InstrumentsJournal of Endodontics, 2012
- Canal preparation using only one Ni‐Ti rotary instrument: preliminary observationsInternational Endodontic Journal, 2007
- Current Challenges and Concepts in the Preparation of Root Canal Systems: A ReviewJournal of Endodontics, 2004
- Defects in Rotary Nickel-Titanium Files After Clinical UseJournal of Endodontics, 2000