Ultra-high vacuum compatible full metal atom beam shutter for optical lattice clocks
- 1 January 2023
- journal article
- conference paper
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 94 (1) , 013201
Abstract
We developed a shutter driven by a solenoid to switch on/off the atomic beam of optical lattice clocks developed at KRISS [C. Y. Park et al., Metrologia 50, 119 (2013), S. Lee et al., New J. Phys. 18, 033030 (2016), H. Kim et al., Jpn. J. Appl. Phys. 56, 050302 (2017), and H. Kim et al., Metrologia 58, 055007 (2021)]. The shutter design was focused on long lifetime and compatibility with an ultra-high vacuum (UHV) environment. Thus, the solenoid was designed to be easily installed and removed from the air-side of a CF flange of the shutter. The flag in the vacuum-side moves only with the simple spring action of a sheet of a metal plate without any frictional movement of mechanical parts. All parts in the vacuum-side were made of metals (stainless steel and pure iron) to be baked over the temperature of 200 °C for UHV. The flag head of the shutter displaces up to 10 mm (5 mm) with a response time of 50 (30 ms) and 80 ms (10 ms) for the opening-action and the closing-action, respectively. The lifetime was tested up to 6 × 106 cycles with no performance degradation. We expect the actual lifetime to be much longer than this by virtue of its friction-free design.Keywords
Funding Information
- Korea Research Institute of Standards and Science (KRISS – 2022 – GP2022-0001)
This publication has 12 references indexed in Scilit:
- Atomic clock performance enabling geodesy below the centimetre levelNature, 2018
- Improved absolute frequency measurement of the171Yb optical lattice clock at KRISS relative to the SI secondJapanese Journal of Applied Physics, 2017
- Atom shutter using bender piezoactuatorReview of Scientific Instruments, 2017
- Cancellation of collisional frequency shifts in optical lattice clocks with Rabi spectroscopyNew Journal of Physics, 2016
- An optical lattice clock with accuracy and stability at the 10−18 levelNature, 2014
- Absolute frequency measurement of1S0(F= 1/2)–3P0(F= 1/2) transition of171Yb atoms in a one-dimensional optical lattice at KRISSMetrologia, 2013
- A Zeeman slower design with permanent magnets in a Halbach configurationReview of Scientific Instruments, 2011
- A Double Magneto-optical Trap of 7Li Loaded by a ??--type Zeeman SlowerJournal of the Korean Physical Society, 2011
- An optical lattice clockNature, 2005