PET Image Quality Improvement for Simultaneous PET/MRI with a Lightweight MRI Surface Coil
- 3 November 2020
- journal article
- research article
- Published by Radiological Society of North America (RSNA) in Radiology
- Vol. 298 (1) , 166-172
Abstract
Background During simultaneous PET/MRI, flexible MRI surface coils that lay on the patient are often omitted from PET attenuation correction processing, leading to quantification bias in PET images. Purpose To identify potential PET image quality improvement by using a recently developed lightweight MRI coil technology for the anterior array (AA) surface coil in both a phantom and in vivo study. Materials and Methods A phantom study and a prospective in vivo study were performed with a PET/CT scanner under three conditions: (a) no MRI surface coil (standard of reference), (b) traditional AA coil, and (c) lightweight AA coil. AA coils were not used in attenuation correction processing to emulate clinical PET/MRI. For the phantom study, PET images were reconstructed with and without time of flight (TOF) to assess quantification accuracy and uniformity. The in vivo study consisted of 10 participants (mean age, 66 years ± 10 [standard deviation]; six men) referred for a PET/CT oncologic examination who had undergone imaging between October 2019 and February 2020. Assessment of image quantification bias (defined as the standard error of the mean values) was conducted by comparing mean liver region of interest standardized uptake values with the no-coil standard of reference. A Wilcoxon signed-rank test was used to establish significance. Results For TOF and non-TOF, respectively, the phantom study revealed a mean PET quantification bias of -9.0% and -8.6% with the traditional AA coil and a mean PET quantification bias of -4.3% and -4.0% with the lightweight AA coil. The coefficients of variation reduced from 4.3% and 6.2% with the traditional AA coil to 2.1% and 2.7% with the lightweight AA coil, which demonstrated a homogeneity benefit from the lightweight coil that was greater with, versus without, TOF reconstruction. For the in vivo study, the mean liver standardized uptake value error was -5.9% with the traditional AA coil (P = .002 vs no coil) and -2.4% with the lightweight AA coil (P = .004 vs no coil). Conclusion The lightweight anterior array coil reduced PET image quantification bias by more than 50% compared with the traditional coil. Using the lightweight coil and performing time of flight-based reconstruction each reduced the variation of error. © RSNA, 2020 Online supplemental material is available for this article.Keywords
This publication has 21 references indexed in Scilit:
- Attenuation Correction for Flexible Magnetic Resonance Coils in Combined Magnetic Resonance/Positron Emission Tomography ImagingInvestigative Radiology, 2014
- Integrated PET/MR imaging: Automatic attenuation correction of flexible RF coilsMedical Physics, 2013
- Simultaneous PET/MR imaging: MR-based attenuation correction of local radiofrequency surface coilsMedical Physics, 2012
- Effects of MR surface coils on PET quantificationMedical Physics, 2011
- The effect of MR surface coils on PET quantification in whole-body PET/MR: Results from a pseudo-PET/MR phantom studyMedical Physics, 2011
- Evaluation of the attenuation properties of MR equipment for its use in a whole-body PET/MR scannerPhysics in Medicine & Biology, 2010
- From RECIST to PERCIST: Evolving Considerations for PET Response Criteria in Solid TumorsJournal of Nuclear Medicine, 2009
- Standards for PET Image Acquisition and Quantitative Data AnalysisJournal of Nuclear Medicine, 2009
- User-guided 3D active contour segmentation of anatomical structures: Significantly improved efficiency and reliabilityNeuroImage, 2006
- Accelerated image reconstruction using ordered subsets of projection dataIEEE Transactions on Medical Imaging, 1994