Abstract
Purpose: To investigate the optical performance of a large-stroke deformable mirror in correcting large aberrations in highly aberrated eyes. Methods: A large-stroke deformable mirror (Mirao 52D; Imagine Eyes) and a Shack-Hartmann wavefront sensor were used in an adaptive optics system. Closed-loop correction of the static aberrations of a phase plate designed for an advanced keratoconic eye was performed for a 6-mm pupil. The same adaptive optics system was also used to correct the aberrations in one eye each of two moderate keratoconic and three normal human eyes for a 6-mm pupil. Results: With closed-loop correction of the phase plate, the total root-mean-square (RMS) over a 6-mm pupil was reduced from 3.54 to 0.04 µm in 30 to 40 iterations, corresponding to 3 to 4 seconds. Adaptive optics closed-loop correction reduced an average total RMS of 1.73±0.998 to 0.10±0.017 µm (higher order RMS of 0.39±0.124 to 0.06±0.004 µm) in the three normal eyes and 2.73±1.754 to 0.10±0.001 µm (higher order RMS of 1.82±1.058 to 0.05±0.017 µm) in the two keratoconic eyes. Conclusions: Aberrations in both normal and highly aberrated eyes were successfully corrected using the large-stroke deformable mirror to provide almost perfect optical quality. This mirror can be a powerful tool to assess the limit of visual performance achievable after correcting the aberrations, especially in eyes with abnormal corneal profiles. [ J Refract Surg. 2007;23:947–952.]