Flip angle optimization for dynamic contrast-enhanced MRI-studies with spoiled gradient echo pulse sequences
Authors:
- D De NaeyerDND De Naeyer
- J VerhulstJVJ Verhulst
- W CeelenW Ceelen
- P SegersPSP Segers
- Y De DeeneYDY De Deene
- P VerdonckPVP Verdonck
- 29 July 2011
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 56 (16) , 5373-5395
Abstract
Spoiled gradient echo pulse (SPGRE) sequences are commonly used in dynamic contrast-enhanced MRI (DCE-MRI) studies to measure the contrast agent concentration in a tissue of interest over time. However, due to improper tuning of the SPGRE parameters, concentration uncertainty can be very high, even at high signal-to-noise ratio in the MR measurement. In this work, an optimization procedure is proposed for selecting the optimal value of the SPGRE-flip angle FA(opt), given the expected concentration range. The optimization condition ensures that every concentration in the assumed range has the lowest possible uncertainty. By decoupling the R(1)- and R*(2)-effects caused by the presence of the contrast agent, a contour plot has been generated from which FA(opt) can be read off for any study design. Investigation of ten recent DCE-MRI studies showed that improper flip angle selection unnecessarily increases the concentration uncertainty, up to 742% and 72% on average for the typical physiological concentration ranges of 0-2 mM in tumour tissue and 0-10 mM in blood, respectively. Simulations show that the reduced noise levels on the concentration curves, observed at the optimal flip angle, effectively increase the precision of the kinetic parameters estimates (up to 82% for K(trans), 82% for ν(e) and 92% for ν(p) in the case of an individually measured arterial input function (AIF), up to 53% for K(trans), 59% for ν(e) and 67% for ν(p) in the case of a standard AIF). In vivo experiments confirm the potential of flip angle optimization to increase the reproducibility of the kinetic parameter estimates.Keywords
This publication has 35 references indexed in Scilit:
- Effects of AZD2171 and vandetanib (ZD6474, Zactima) on haemodynamic variables in an SW620 human colon tumour model: an investigation using dynamic contrast‐enhanced MRI and the rapid clearance blood pool contrast agent, P792 (gadomelitol)NMR in Biomedicine, 2007
- AZD2171, a Pan-VEGF Receptor Tyrosine Kinase Inhibitor, Normalizes Tumor Vasculature and Alleviates Edema in Glioblastoma PatientsCancer Cell, 2007
- Noninvasive monitoring of radiotherapy-induced microvascular changes using dynamic contrast enhanced magnetic resonance imaging (DCE-MRI) in a colorectal tumor modelInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2006
- Rapid high‐resolution T1 mapping by variable flip angles: Accurate and precise measurements in the presence of radiofrequency field inhomogeneityMagnetic Resonance in Medicine, 2006
- Prostate Cancer: Evaluation of Vascular Characteristics with Dynamic Contrast-enhanced T1-weighted MR Imaging—Initial ExperienceRadiology, 2004
- Dynamic magnetic resonance imaging of tumor perfusionIEEE Engineering in Medicine and Biology Magazine, 2004
- BASIC PULSE SEQUENCES2004
- Key factors in the acquisition of contrast kinetic data for oncologyJournal of Magnetic Resonance Imaging, 1999
- Accuracy of T1 measurement in dynamic contrast-enhanced breast MRI using two- and three-dimensional variable flip angle fast low-angle shotJournal of Magnetic Resonance Imaging, 1999
- Contrast in Rapid MR ImagingJournal of Computer Assisted Tomography, 1987