A Sub-mm$^3$ Ultrasonic Free-floating Implant for Multi-mote Neural Recording
Authors:
- Mohammad Meraj GhanbariMGMohammad Meraj Ghanbari
- David K. PiechDPDavid K. Piech
- Konlin ShenKSKonlin Shen
- Sina Faraji AlamoutiSASina Faraji Alamouti
- Cem YalcinCYCem Yalcin
- Benjamin C. JohnsonBJBenjamin C. Johnson
- Jose M. CarmenaJCJose M. Carmena
- Michel M. MaharbizMMMichel M. Maharbiz
- Rikky MullerRMRikky Muller
Preprint
Abstract
A 0.8 mm$^3$ wireless, ultrasonically powered, free-floating neural recording implant is presented. The device is comprised only of a 0.25 mm$^2$ recording IC and a single piezoceramic resonator that is used for both power harvesting and data transmission. Uplink data transmission is performed by analog amplitude modulation of the ultrasound echo. Using a 1.78 MHz main carrier, >35 kbps/mote equivalent uplink data rate is achieved. A technique to linearize the echo amplitude modulation is introduced, resulting in 2.5x, resulting in the highest measured depth/volume ratio by $\sim$3x. Orthogonal subcarrier modulation enables simultaneous operation of multiple implants, using a single-element ultrasound external transducer. Dual-mote simultaneous power up and data transmission is demonstrated at a rate of 7 kS/s at the depth of 50 mm.Keywords
All Related Versions
This publication has 18 references indexed in Scilit:
- Cervical vagus nerve morphometry and vascularity in the context of nerve stimulation - A cadaveric studyScientific Reports, 2018
- A Miniaturized Single-Transducer Implantable Pressure Sensor With Time-Multiplexed Ultrasonic Data and Power LinksIEEE Journal of Solid-State Circuits, 2018
- Comparison of the in-vivo neural recording quality of floating and skull-fixed silicon probes2017
- Reliable Next-Generation Cortical Interfaces for Chronic Brain–Machine Interfaces and NeuroscienceProceedings of the IEEE, 2016
- Silicon-Integrated High-Density Electrocortical InterfacesProceedings of the IEEE, 2016
- Wireless Recording in the Peripheral Nervous System with Ultrasonic Neural DustNeuron, 2016
- Chronic tissue response to untethered microelectrode implants in the rat brain and spinal cordJournal of Neural Engineering, 2015
- A Minimally Invasive 64-Channel Wireless μECoG ImplantIEEE Journal of Solid-State Circuits, 2014
- Two novel fully complementary self-biased CMOS differential amplifiersIEEE Journal of Solid-State Circuits, 1991
- New equivalent circuits for elementary piezoelectric transducersElectronics Letters, 1970