Contemporary femoral designs in total knee arthroplasty: effects on the patello-femoral congruence
- 28 December 2011
- journal article
- research article
- Published by Springer Nature in International Orthopaedics
- Vol. 36 (6) , 1167-1173
Abstract
The objective of this study was to evaluate the radiological and clinical correlations between implant design and patellar positioning in patients who underwent TKA utilizing femoral implants with modern designs. Thirty consecutive PFC PS Sigma TKAs, characterized by a new prolonged anterior flange and a “smoother” trochlea, were prospectively reviewed. All patellae were replaced. All patients were evaluated pre-operatively and prospectively at two years follow-up both clinically according to the Knee Society score as well as radiographically. This included computed tomography (CT); patellar tilt, patellar conformity angle, patellar lateralization, and femoral component external-rotation in relation to the clinical trans-epicondylar axis. Average patellar tilt at follow-up was 3° (±7.5°) with respect to a pre-operative 18.5° (±8.5°). Average patellar congruence angle at follow-up was −3° (range, −11° to +9°) with respect to a pre-operative 10.3° (range, + 1.5° to 25.5°). Average lateralization index at follow-up was 2.7 mm (range, −3.4 mm to +7.1 mm) with respect to a pre-operative 12.2 mm (± 4.8 mm). Femoral component positioning related to the trans-epicondylar axis showed an external rotation of 2.80° (± 2.10°) at follow-up with respect to 5.7° (± 1.80°). Clinically, two (6.6%) patients reported patello-femoral complications related to imperfections in the surgical technique more than the implant's design. This study highlighted that modern femoral designs in TKA allow for a correct reproducibility of a normal patello-femoral conformity. Strict surgical principles are paramount to avoid patello-femoral complications even when modern implants are used.Keywords
This publication has 48 references indexed in Scilit:
- What is the Evidence for Total Knee Arthroplasty in Young Patients?: A Systematic Review of the LiteratureClinical Orthopaedics and Related Research, 2011
- Systematic review of patellar resurfacing in total knee arthroplastyInternational Orthopaedics, 2010
- Analysis of the outcome in male and female patients using a unisex total knee replacement systemThe Journal of Bone and Joint Surgery. British volume, 2009
- Rotational Position of Femoral and Tibial Components in TKA Using the Femoral Transepicondylar AxisClinical Orthopaedics and Related Research, 2008
- The trochlea is medialized by total knee arthroplasty: An intraoperative assessment in 61 patientsActa Orthopaedica, 2007
- Magnetic resonance imaging of in vivo kinematics after total knee arthroplastyJournal of Magnetic Resonance Imaging, 2005
- Rotating Platform Knees Did Not Improve Patellar TrackingClinical Orthopaedics and Related Research, 2004
- Reliability of the anteroposterior axis and the posterior condylar axis for determining rotational alignment of the femoral component in total knee arthroplastyJournal of Orthopaedic Science, 1998
- Patellar motion analyzed by magnetic resonance imagingActa Orthopaedica, 1989
- A New Patella ProsthesisClinical Orthopaedics and Related Research, 1975