The influence of exposure parameters on jawbone model accuracy using cone beam CT and multislice CT

Authors:

Abstract
Objective: The main purpose of this study was to investigate the influence of exposure parameters on jawbone model accuracy when using cone beam CT (CBCT) and multislice CT (MSCT). Methods: A lower and an upper edentulous human cadaver jaw were scanned using micro-CT (Skyscan 1173 high energy spiral scan micro-CT; Skyscan NV, Kontich, Belgium) at 35 μm to serve as true reference. The in vitro samples were exposed using six CBCT units and one MSCT system. CBCT exposure protocols were chosen according to clinically available settings. The variables were kilovoltage, milliamperage, voxel size and/or scan time. Image segmentation was based on local thresholds using profile lines. The resulting jawbone segmentations were registered with the reference and image processing was done to internally fill the segmentations. A point-based distance calculation was performed between the three-dimensional objects and reference scans and deviation percentages were calculated for 2 mm, 1 mm and 0.5 mm intervals. Results: A...