Comparative clinical study of canine and feline total blood cell count results with seven in‐clinic and two commercial laboratory hematology analyzers
Authors:
- 19 November 2008
- journal article
- Published by Wiley in Veterinary Clinical Pathology
- Vol. 37 (4) , 373-384
Abstract
Background: A CBC is an integral part of the assessment of health and disease in companion animals. While in the past newer technologies for CBC analysis were limited to large clinical pathology laboratories, several smaller and affordable automated hematology analyzers have been developed for in‐clinic use.Objectives: The purpose of this study was to compare CBC results generated by 7 in‐clinic laser‐ and impedance‐based hematology instruments and 2 commercial laboratory analyzers.Methods: Over a 3‐month period, fresh EDTA‐anticoagulated blood samples from healthy and diseased dogs (n=260) and cats (n=110) were analyzed on the LaserCyte, ForCyte, MS45, Heska CBC, Scil Vet ABC, VetScan HMT, QBC Vet Autoread, CELL‐DYN 3500, and ADVIA 120 analyzers. Results were compared by regression correlation (linear, Deming, Passing‐Bablok) and Bland–Altman bias plots using the ADVIA as the criterion standard for all analytes except HCT, which was compared with manual PCV. Precision, linearity, and carryover also were evaluated.Results: For most analytes, the in‐clinic analyzers and the CELL‐DYN performed similarly and correlated well with the ADVIA. The biases ranged from −0.6 to 2.4 × 109/L for WBC count, 0 to 0.9 × 1012/L for RBC count, −1.5 to 0.7 g/dL for hemoglobin concentration, −4.3 to 8.3 fL for MCV, and −69.3 to 77.2 × 109/L for platelet count. Compared with PCV, the HCT on most analyzers had a bias from 0.1% to 7.2%. Canine reticulocyte counts on the LaserCyte and ForCyte correlated but had a negative bias compared with those on the ADVIA. Precision, linearity, and carryover results were excellent for most analyzers.Conclusions: Total WBC and RBC counts were acceptable on all in‐clinic hematology instruments studied, with limitations for some RBC parameters and platelet counts. Together with evaluation of a blood film, these in‐clinic instruments can provide useful information on canine and feline patients in veterinary practices.Keywords
This publication has 16 references indexed in Scilit:
- Clinical evaluation of the CA530‐VET hematology analyzer for use in veterinary practiceVeterinary Clinical Pathology, 2007
- Comparison of white blood cell differential percentages determined by the in‐house LaserCyte hematology analyzer and a manual methodVeterinary Clinical Pathology, 2006
- Analysis of canine and feline haemograms using the VetScan HMT analyserJournal of Small Animal Practice, 2003
- Artifactual Changes in Equine Blood Following Storage, Detected Using the Advia 120 Hematology AnalyzerVeterinary Clinical Pathology, 2002
- Evaluation of methods of platelet counting in the catJournal of Small Animal Practice, 2001
- Evaluation of an in-house centrifugal hematology analyzer for use in veterinary practiceJournal of the American Veterinary Medical Association, 2000
- Analysis of Feline, Canine and Equine Hemograms Using the QBC VetAutoreadVeterinary Clinical Pathology, 1999
- Evaluation of the QBC-vet autoread haematology system for domestic and pet animal speciesComparative Clinical Pathology, 1998
- Comparing Laboratory Tests Using the Difference Plot MethodVeterinary Clinical Pathology, 1993
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986