Shunt Dysfunction Assessment in Shunted Patients via Multiparametric Non-Invasive ICP Waveform Monitoring
Authors:
- Raphael BertaniRBRaphael Bertani
- Gabriel SemioneGSGabriel Semione
- Milene Zanella CapitanioMCMilene Zanella Capitanio
- Sérgio BrasilSBSérgio Brasil
- Sávio BatistaSBSávio Batista
- Gabriel André da Silva MendesGMGabriel André da Silva Mendes
- Caio PerretCPCaio Perret
- Christian FerreiraCFChristian Ferreira
- Wellingson Silva PaivaWPWellingson Silva Paiva
- Fernando Campos Gomes PintoFPFernando Campos Gomes Pinto
Open Access
- 7 October 2025
- journal article
- research article
- Published by MDPI AG in Biomedicines
- Vol. 13 (10) , 2436
Abstract
Background/Objectives: Ventriculoperitoneal shunts are the standard treatment for adults diagnosed with several CSF disorders, but often face dysfunction, leading interest in non-invasive methods for diagnosing shunt issues. This study evaluates the potential of non-invasive intracranial pressure waveform monitoring (nICPw) with the brain4care (B4C) system to distinguish overdrainage, underdrainage, and normal shunt function in patients with CSF disorders. Methods: In this single-center, observational study at Hospital das Clínicas, Brazil, adult patients with CSF shunts were enrolled. Patients were categorized as Overdrainage or Underdrainage, based on clinical parameters, with an Asymptomatic group. The B4C system provided nICPw monitoring, and six parameters (including various P2/P1 ratios) were analyzed via MANOVA and ANOVA. Results: Among 30 patients (6 overdrainage, 6 underdrainage, 18 asymptomatic), five asymptomatic patients were excluded from the main analysis due to incomplete data collection, leaving 25 patients. Overdrainage patients had significantly higher ΔP2/P1 values (0.618 ± 0.210) than asymptomatic ones (0.227 ± 0.171). After excluding outliers, differences were more pronounced (H = 10.89; p < 0.01). Underdrainage patients had intermediate ΔP2/P1 values (0.387 ± 0.179) and consistently higher P2/P1 averages (>1.3). ROC analysis indicated that ΔP2/P1 > 0.3 suggested shunt dysfunction (AUC = 0.731), while the highest P2/P1 offered stronger discrimination (AUC = 0.782). A global average P2/P1 > 1.3 was linked to underdrainage, with the lowest P2/P1 values differentiating overdrainage (0.948 ± 0.321) from underdrainage (1.143 ± 0.156). Conclusions: nICPw monitoring with the B4C system demonstrated potential for detecting shunt dysfunction. Combining parameters, especially ΔP2/P1 and highest P2/P1, improves diagnostic accuracy, offering a non-invasive method that may aid in distinguishing normal from abnormal shunt function.Keywords
This publication has 28 references indexed in Scilit:
- Validation of a Noninvasive Approach for Cerebrospinal Compliance MonitoringNeurocritical Care, 2025
- Noninvasive Method Using Mechanical Extensometer for the Estimation of Intracranial Compliance by Repeated Measures Agreement AnalysisNeurocritical Care, 2024
- Noninvasive intracranial pressure monitoring throughout brain compliance guiding a ventriculoperitoneal shunt replacement in hydrocephalus—case reportChild's Nervous System, 2023
- Intracranial pressure: current perspectives on physiology and monitoringIntensive Care Medicine, 2022
- Waveform Morphology as a Surrogate for ICP Monitoring: A Comparison Between an Invasive and a Noninvasive MethodNeurocritical Care, 2022
- Comparison of Waveforms Between Noninvasive and Invasive Monitoring of Intracranial Pressure2021
- Overdrainage of cerebrospinal fluid and hydrocephalus shuntsActa Neurochirurgica, 2017
- Non-invasive Monitoring of Intracranial Pressure Using Transcranial Doppler Ultrasonography: Is It Possible?Neurocritical Care, 2016
- Transcranial Doppler sonography in adult hydrocephalic patientsNeurosurgical Review, 2000
- Value of transcranial Doppler indices in predicting raised ICP in infantile hydrocephalusChild's Nervous System, 1995