A Concept for Magnetic Resonance Visualization of Surgical Textile Implants
Authors:
- Nils A. KrämerNKNils A. Krämer
- Hank C. W. DonkerHDHank C. W. Donker
- Jens OttoJOJens Otto
- Michael HodeniusMHMichael Hodenius
- Julien SénégasJSJulien Sénégas
- Ioana SlabuISIoana Slabu
- Uwe KlingeUKUwe Klinge
- Martin BaumannMBMartin Baumann
- Andreas MüllenAMAndreas Müllen
- Boris ObolenskiBOBoris Obolenski
- Rolf W. GüntherRGRolf W. Günther
- Gabriele A. KrombachGKGabriele A. Krombach
- 1 August 2010
- journal article
- Published by Wolters Kluwer Health in Investigative Radiology
- Vol. 45 (8) , 477-483
Abstract
To develop a method for visualizing surgical textile implant (STI) with superparamagnetic iron oxides (SPIO), using magnetic resonance imaging (MRI). Therefore, positive-contrast inversion-recovery with on-resonant water suppression (IRON) was applied and its properties were evaluated in vitro. STI with different concentrations of SPIO integrated into the base material were produced. Imaging was performed on a clinical 1.5 Tesla scanner, using conventional balanced gradient echo sequences (SSFP), T2*-weighted sequences, and IRON-imaging. In vitro experiments were conducted in an agarose phantom. On MR-images, contrast-to-noise-ratios, and the dimensions of the implant were assessed. Conventional MRI exhibited SPIO-loaded STI as signal voids. Using IRON, the mesh was clearly exhibited hyperintensely with suppression of on-resonant background signals with a distinct differentiation to other sources of off-resonances. Concentrations of approximately 9 mg/g led to best positive contrast and highest contrast-to-noise-ratios using IRON. Depending on B0-orientation, phase encoding direction and the STI's SPIO-load, the IRON-signal showed a characteristic pattern and an overestimation of STI size up to 4.6 mm. The integration of SPIOs into the base material combined with IRON is a feasible approach to visualize STI with MRI. This method could help to identify mesh-related problems in time and to reduce the need for surgical revision.Keywords
This publication has 33 references indexed in Scilit:
- Positive contrast visualization for cellular magnetic resonance imaging using susceptibility-weighted echo-time encodingMagnetic Resonance Imaging, 2009
- Off-Resonance Angiography: A New Method to Depict Vessels—Phantom and Rabbit StudiesRadiology, 2008
- Positive contrast visualization of nitinol devices using susceptibility gradient mappingMagnetic Resonance in Medicine, 2008
- Susceptibility gradient mapping (SGM): A new postprocessing method for positive contrast generation applied to superparamagnetic iron oxide particle (SPIO)‐labeled cellsMagnetic Resonance in Medicine, 2008
- Noninvasive Detection of Macrophage-Rich Atherosclerotic Plaque in Hyperlipidemic Rabbits Using “Positive Contrast” Magnetic Resonance ImagingJACC, 2008
- Positive contrast MR‐lymphography using inversion recovery with ON‐resonant water suppression (IRON)Journal of Magnetic Resonance Imaging, 2008
- Mesh repair of common abdominal hernias: a review on experimental and clinical studiesHernia, 2008
- Chronic Pain after Inguinal HerniorrhaphyJournal of the American College of Surgeons, 2007
- MR contrast agents: Physical and pharmacologic basicsJournal of Magnetic Resonance Imaging, 2007
- The Management of Incisional HerniaThe Annals of The Royal College of Surgeons of England, 2006