A Concept for Magnetic Resonance Visualization of Surgical Textile Implants

Authors:
  • HDHank C. W. Donker
    Hank C. W. Donker
  • JOJens Otto
    Jens Otto
  • MHMichael Hodenius
    Michael Hodenius
  • JSJulien Sénégas
    Julien Sénégas
  • UKUwe Klinge
    Uwe Klinge
  • MBMartin Baumann
    Martin Baumann
  • AMAndreas Müllen
    Andreas Müllen
  • BOBoris Obolenski
    Boris Obolenski
  • RGRolf W. Günther
    Rolf W. Günther
  • GKGabriele A. Krombach
    Gabriele A. Krombach

Abstract
To develop a method for visualizing surgical textile implant (STI) with superparamagnetic iron oxides (SPIO), using magnetic resonance imaging (MRI). Therefore, positive-contrast inversion-recovery with on-resonant water suppression (IRON) was applied and its properties were evaluated in vitro. STI with different concentrations of SPIO integrated into the base material were produced. Imaging was performed on a clinical 1.5 Tesla scanner, using conventional balanced gradient echo sequences (SSFP), T2*-weighted sequences, and IRON-imaging. In vitro experiments were conducted in an agarose phantom. On MR-images, contrast-to-noise-ratios, and the dimensions of the implant were assessed. Conventional MRI exhibited SPIO-loaded STI as signal voids. Using IRON, the mesh was clearly exhibited hyperintensely with suppression of on-resonant background signals with a distinct differentiation to other sources of off-resonances. Concentrations of approximately 9 mg/g led to best positive contrast and highest contrast-to-noise-ratios using IRON. Depending on B0-orientation, phase encoding direction and the STI's SPIO-load, the IRON-signal showed a characteristic pattern and an overestimation of STI size up to 4.6 mm. The integration of SPIOs into the base material combined with IRON is a feasible approach to visualize STI with MRI. This method could help to identify mesh-related problems in time and to reduce the need for surgical revision.

This publication has 33 references indexed in Scilit: