Interactively variable isotropic resolution in computed tomography
Authors:
- Yiannis KyriakouYKYiannis Kyriakou
- Marc KachelrießMKMarc Kachelrieß
- Sylvia WilharmSWSylvia Wilharm
- Willi A KalenderWKWilli A Kalender
- 30 April 2008
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 53 (10) , 2693-2713
Abstract
An individual balancing between spatial resolution and image noise is necessary to fulfil the diagnostic requirements in medical CT imaging. In order to change influencing parameters, such as reconstruction kernel or effective slice thickness, additional raw-data-dependent image reconstructions have to be performed. Therefore, the noise versus resolution trade-off is time consuming and not interactively applicable. Furthermore, isotropic resolution, expressed by an equivalent point spread function (PSF) in every spatial direction, is important for the undistorted visualization and quantitative evaluation of small structures independent of the viewing plane. Theoretically, isotropic resolution can be obtained by matching the in-plane and through-plane resolution with the aforementioned parameters. Practically, however, the user is not assisted in doing so by current reconstruction systems and therefore isotropic resolution is not commonly achieved, in particular not at the desired resolution level. In this paper, an integrated approach is presented for equalizing the in-plane and through-plane spatial resolution by image filtering. The required filter kernels are calculated from previously measured PSFs in x/y- and z-direction. The concepts derived are combined with a variable resolution filtering technique. Both approaches are independent of CT raw data and operate only on reconstructed images which allows for their application in real time. Thereby, the aim of interactively variable, isotropic resolution is achieved. Results were evaluated quantitatively by measuring PSFs and image noise, and qualitatively by comparing the images to direct reconstructions regarded as the gold standard. Filtered images matched direct reconstructions with arbitrary reconstruction kernels with standard deviations in difference images of typically between 1 and 17 HU. Isotropic resolution was achieved within 5% of the selected resolution level. Processing times of 20-100 ms per frame allow for interactive use.Keywords
This publication has 10 references indexed in Scilit:
- Presampling, algorithm factors, and noise: Considerations for CT in particular and for medical imaging in generalMedical Physics, 2005
- Spatial domain filtering for fast modification of the tradeoff between image sharpness and pixel noise in computed tomographyIEEE Transactions on Medical Imaging, 2003
- Spatial domain image filtering in computed tomography: feasibility study in pulmonary embolismEuropean Radiology, 2003
- Fast image filters as an alternative to reconstruction kernels in computed tomography2001
- Novel approximate approach for high-quality image reconstruction in helical cone-beam CT at arbitrary pitch2001
- Effect of scattered radiation on image noise in cone beam CTMedical Physics, 2001
- Advanced single‐slice rebinning in cone‐beam spiral CTMedical Physics, 2000
- Selection of task-dependent diffusion filters for the post-processing of SPECT imagesPhysics in Medicine & Biology, 1998
- Structure adaptive anisotropic image filteringImage and Vision Computing, 1996
- Radiologic image compression-a reviewProceedings of the IEEE, 1995