Experimental verification of the applicability of the AQURE IOERT accelerator for FLASH radiotherapy
Authors:
- Aleksandra Lenartowicz-GasikALAleksandra Lenartowicz-Gasik
- Agnieszka MisiarzAMAgnieszka Misiarz
- Przemysław AdrichPAPrzemysław Adrich
- Sławomir WronkaSWSławomir Wronka
- Tomasz PiotrowskiTPTomasz Piotrowski
- Bartosz PawałowskiBPBartosz Pawałowski
- Piotr NowaczykPNPiotr Nowaczyk
- Jan TrzuskowskiJTJan Trzuskowski
- Jacek RzadkiewiczJacek Rzadkiewicz
Open Access
- 1 March 2025
- journal article
- Published by Walter de Gruyter GmbH in Polish Journal of Medical Physics And Engineering
- Vol. 31 (1) , 62-72
Abstract
Introduction: FLASH radiotherapy is based on delivering an ultra-high dose of ionizing radiation in a very short time, which allows for the enhanced protection of healthy tissues and the effective destruction of tumour tissues. The paper presents experimental verification of the applicability of the intraoperative electron radiotherapy (IOERT) accelerator to obtain electron beams with ultra-high dose rates. The results of electron beam dosimetric measurements were compared with simulations to verify the parameters of the electron beams obtained in FLASH mode. Material and Methods: The IOERT AQURE accelerator is designed and manufactured at the National Centre for Nuclear Research. The device provides electron beams with energies in the energy range of 4-12 MeV. The accelerator was recently upgraded to deliver ultra-high-dose rates (FLASH). The emitted electron beams were verified by percentage dose depth (PDD) and beam profile measurements using Gafchromic EBT-XD films. The measurement results were compared with Monte Carlo simulations performed using the Geant4 application. Results: Conventional and FLASH electron beam parameters, i.e. depths of doses, most probable beam energies at the phantom surface, and average beam energies at the phantom surface were determined experimentally for the upgraded IOERT accelerator for 6 MeV and 9 MeV. The results of dose measurements showed that the accelerator can provide electron beams with homogeneous and symmetrical fields with dose rates up to 300 Gy/s ± 6 Gy/s, for an electron beam at a nominal energy of 6 MeV, and up to 440 Gy/s ± 9 Gy/s at a nominal energy of 9 MeV. A comparison of measured and simulated PDDs and dose profiles for FLASH beams showed good agreement (no worse than 90%) under the gamma index of 3%/3 mm. Conclusions: The results obtained confirm that the upgraded intraoperative accelerator can be used in planned, pre-clinical radiobiology experiments in both conventional and ultra-high dose rate (FLASH-RT) modes.Keywords
This publication has 22 references indexed in Scilit:
- Ultra‐high dose rate dosimetry: Challenges and opportunities for FLASH radiation therapyMedical Physics, 2022
- Abdominal FLASH irradiation reduces radiation-induced gastrointestinal toxicity for the treatment of ovarian cancer in miceScientific Reports, 2020
- Treatment of a first patient with FLASH-radiotherapyRadiotherapy and Oncology, 2019
- Technical Note: Monte Carlo study on the reduction in x‐ray contamination of therapeutic electron beams for Intraoperative Radiation Therapy by means of improvements in the design of scattering foilsMedical Physics, 2019
- Technical Note: Time‐gating to medical linear accelerator pulses: Stray radiation detectorMedical Physics, 2018
- High dose‐per‐pulse electron beam dosimetry — A model to correct for the ion recombination in the Advanced Markus ionization chamberMedical Physics, 2017
- Dose rate dependence for different dosimeters and detectors: TLD, OSL, EBT films, and diamond detectorsMedical Physics, 2012
- Radiation Oncology Physics: A Handbook for Teachers and StudentsBritish Journal of Cancer, 2008
- Applications of the General Particle Tracer code2002
- A technique for the quantitative evaluation of dose distributionsMedical Physics, 1998