    3D printing from a biocompatible material for a new design of electron applicators of the intraoperative accelerator

    by Agnieszka Misiarz, Łukasz Kujawiński, Marek Baran, Michał Kaczmarek, Jan Trzuskowski, Karina Guzdek, Marta Kruszyna-Mchalska, Bartosz Pawałowski

    Introduction: A special type of radiotherapy is Intraoperative Electron Radiation Therapy (IOERT). The electron applicator is the element forming the final shape of the electron beam delivered to the postoperative bed. This work presents the new design of an electron applicator made from the biocompatible material MED610. The applicators will form an electron beam in the AQURE intraoperative accelerator. Material and methods: The main problem to be solved during the construction work was to propose a design solution for the electron applicator that would be light, transparent, and made of a biocompatible material, and at the same time sufficiently shield the patient against radiation scattered outside its walls. Moreover, the applicator used in intraoperative therapy should have the thinnest possible walls. Results: The calculations and design work performed allowed the lower part of the applicator to be made of a biocompatible, material. This element allows the beam to be formed in accordance with the normative recommendations, it is light and transparent. In this case, PolyJet 3D printing was selected as the most appropriate method. Conclusions: Thanks to the use of a new solution for the shape of the walls, the applicator limits the leakage radiation by the standard, without the need for thicker outer walls. This material is light, thanks to which the doctor keeps this part of the applicator in the right position until it is fastened with the part on the head that does not require additional, dedicated fastening. The resin MED610 is suitable for 3D printing, it can be easily used to create any shape, and thus the possibilities of forming the lower part of the applicator are not limited. In the future, for individual purposes, the shape of the beam formed by the lower part of the applicator may be dedicated to special applications.

  • Open Access

    Experimental verification of the applicability of the AQURE IOERT accelerator for FLASH radiotherapy

    by Aleksandra Lenartowicz-Gasik, Agnieszka Misiarz, Przemysław Adrich, Sławomir Wronka, Tomasz Piotrowski, Bartosz Pawałowski, Piotr Nowaczyk, Jan Trzuskowski, Jacek Rzadkiewicz

    Introduction: FLASH radiotherapy is based on delivering an ultra-high dose of ionizing radiation in a very short time, which allows for the enhanced protection of healthy tissues and the effective destruction of tumour tissues. The paper presents experimental verification of the applicability of the intraoperative electron radiotherapy (IOERT) accelerator to obtain electron beams with ultra-high dose rates. The results of electron beam dosimetric measurements were compared with simulations to verify the parameters of the electron beams obtained in FLASH mode. Material and Methods: The IOERT AQURE accelerator is designed and manufactured at the National Centre for Nuclear Research. The device provides electron beams with energies in the energy range of 4-12 MeV. The accelerator was recently upgraded to deliver ultra-high-dose rates (FLASH). The emitted electron beams were verified by percentage dose depth (PDD) and beam profile measurements using Gafchromic EBT-XD films. The measurement results were compared with Monte Carlo simulations performed using the Geant4 application. Results: Conventional and FLASH electron beam parameters, i.e. depths of doses, most probable beam energies at the phantom surface, and average beam energies at the phantom surface were determined experimentally for the upgraded IOERT accelerator for 6 MeV and 9 MeV. The results of dose measurements showed that the accelerator can provide electron beams with homogeneous and symmetrical fields with dose rates up to 300 Gy/s ± 6 Gy/s, for an electron beam at a nominal energy of 6 MeV, and up to 440 Gy/s ± 9 Gy/s at a nominal energy of 9 MeV. A comparison of measured and simulated PDDs and dose profiles for FLASH beams showed good agreement (no worse than 90%) under the gamma index of 3%/3 mm. Conclusions: The results obtained confirm that the upgraded intraoperative accelerator can be used in planned, pre-clinical radiobiology experiments in both conventional and ultra-high dose rate (FLASH-RT) modes.


  • Open Access

    Verification of real-time dosimetry of ultra-high dose rate beams at AQURE FLASH RT on the sample surface

    by Aleksandra Lenartowicz-Gasik, Agnieszka Misiarz, Vasyl Markopolski, Marta Walo, Urszula Gryczka, Sylwester Bułka, Bartosz Pawałowski, Marta Kruszyna-Mochalska, Jacek Rzadkiewicz

    Introduction: FLASH radiotherapy involves delivering a dose of ultra-high-dose-rate ionising radiation (>40 Gy/s) in less than 200 ms, resulting in sparing healthy tissue and effectively destroying cancerous tissue. This article presents a preliminary verification of the feasibility of using real-time internal dosimetry at the sample surface to measure doses generated by the AQURE FLASH RT accelerator dedicated to FLASH radiotherapy studies. Material and Methods: The AQURE FLASH RT emits a 6 and 9 MeV electron beam with a dose rate higher than 40 Gy/s. The real-time dosimetry system to measure doses on the sample surface was implemented into the accelerator and consists of inductive sensors in which the moving electron charge induces a voltage in the secondary toroidal winding. The internal dosimetric system was preliminarily calibrated for single pulse irradiations using passive dosimetry methods, i.e. film and alanine dosimetry. Results: The study showed that there was no effect of the tested dosimetry system on the beam (PDD and beam profile). The linearity of the system’s response to successive pulses was tested and found to be Conclusion: The results confirmed the possibility of using the real-time dosimetry system to measure doses on the sample surface delivered by ultra-high dose rate beams at the AQURE FLASH RT accelerator. The system has been tested and validated over the full range of dose rates emitted by the accelerator to measure a dose in a single pulse. The results of the dosimetric measurements confirmed that the system did not affect the beam parameters.


  • Open Access

    Commissioning, dosimetric characterisation and machine performance assessment of the AQURE mobile accelerator for intraoperative radiotherapy

    by Adam Ryczkowski, Bartosz Pawałowski, Marta Kruszyna-Mochalska, Agnieszka Misiarz, Aleksandra Lenartowicz-Gasik, Maksymilian Wosicki, Agata Jodda, Przemysław Adrich, Tomasz Piotrowski

    Introduction: The AQURE is a recently designed mobile linear accelerator for intraoperative electron radiotherapy (IOERT). In the study, we assess, report, and compare the dosimetric characteristics obtained during the commissioning of this machine in our Centre. Material and Methods: The electron energies of the AQURE used in this study are 4, 6, 9 and 12 MeV. The diameters of the cylindrically shaped applicators range from 4 to 10 cm. The measurements were performed (a) by microDiamond detector in a BeamScan water phantom (depth dose distributions, flatness and symmetry of dose profiles, output constancy and linearity) and (b) by Farmer ionisation chamber in a solid water phantom (output related to gantry angle). The results of measurements were compared to analogous results from other IOERT accelerators and were evaluated in the light of scientific recommendations and legal regulations. Results: The values of the measured parameters fit the recommended ranges. The profiles and the depth dose distributions are close to literature data for other IOERT machines. Any differences between them are caused by the detailed technical solutions which influence the final shape of the obtained dose distributions. The values of mean and most probable energies suggest that the electron beams used during IOERT cannot be considered mono-energetic. We believe that the energy spectrum should be researched and described for each IOERT mobile accelerator before clinical usage. Conclusion: The measurements confirm the dosimetric accuracy of the AQURE accelerator under the literature guidelines.


  • Open Access

    Design and performance validation of a novel 3d printed thin-walled and transparent electron beam applicators for intraoperative radiation therapy with beam energy up to 12 MeV

    by Agnieszka Misiarz, Aleksandra Lenartowicz, Przemysław Adrich, Jacek Rzadkiewicz, Sławomir Wronka, Jan Trzuskowski, Marta Kruszyna-Mochalska, Bartosz Urbański, Beata Adamczyk, Jacek Pracz

    A high-energy electron accelerator is used in the treatment of patients in the so-called intraoperative electron radiotherapy (IOERT). The work aimed to present the results of the validation of a new design of an electron beam applicator for use in IOERT. A novel solution was described along with the design optimization method based on Monte Carlo simulations. In this solution, the applicator consists of two parts. The lower exchangeable part collimates the therapeutic field. Measurements were made based on the International Electrotechnical Commission (IEC) standard recommendations. The measurement described in the standard has been adapted to the specificity of the intraoperative accelerator Source to Skin Distance — of 60 cm and applicators with a circular cross-sectional area. Measurements were performed for nominal beam energies of 6, 10, and 12 MeV and two therapeutic field diameters of 6 and 10 cm. The dose due to stray X-ray radiation in all energies is less than 0.3% and increases for energies from 6 to 12 MeV by 2.9 times from 0.1 for 6MeV to 0.29 for 12 MeV. The average dose due to leakage radiation also shows an increasing trend and is higher for a 6 cm diameter applicator. Validation confirmed the usefulness of the novel applicator design for clinical applications. Thanks to the use of 3D printing, it was possible to make applicators that are transparent, biocompatible and, at the same time, light and form a beam field with therapeutically useful accuracy, and the leakage radiation does not exceed normative recommendations. Keywords: IOERT electron linear accelerator electron applicator


  • 2 inch molecular organic glass scintillator for neutron–gamma discrimination

    by M. Grodzicka-Kobylka, T. Szczesniak, M. Moszyński, L. Swiderski, K. Brylew, P.L. Feng, L. Nguyen, J.S. Carlson, J.J. Valiente-Dobón, J. Trzuskowski, et al.


  • Open Access

    Solutions and constructional modifications of dosimetry systems ofelectron accelerators

    by Agnieszka Misiarz, Bożena Malesa, Wasyl Markopolskyi, Ewa Jakubowska, Jerzy Wyspokiński

    The study is designed to present a method of dosimetry system solution in electron accelerators manufactured at the NCBJ Nuclear Equipment Division (HITEC). Two special cases of accelerator applications have been presented - the first using a high-efficiency electron beam in material research and the second using a low-efficiency photon beam in cargo inspection at borders. The article presents constructional modifications of the standardly used ionization chamber, which successfully enabled proper dose measurement in both of the above-mentioned cases.

  • Open Access

    The IOERT IntraLine accelerator – the development, current state, and future plans

    by Przemysław Adrich, Radosław Hanke, Ewa Kulczycka, Konrad Kosiński, Bartosz Meglicki, Agnieszka Misiarz, Eugeniusz Pławski, Marcin Staszczak, Kaja Swat, Agnieszka Syntfeld-Każuch, et al.

    Intraoperative electron radiation therapy (IOERT) is a cancer treatment method that combines oncologic surgery with electron beam radiotherapy. This method can reduce the duration of entire tumor treatment and increase its effectiveness. Moreover, shortening the treatment time significantly reduces the cost and accessibility of the therapy. As a result of interdisciplinary research, an innovative accelerator for IOERT, the IntraLine, was developed. In the course of this work, four patent applications were filed. Today, the work is half way through. In the near future, the device will be optimized. New mechanical design solutions will be developed. Mechanical optimization will significantly reduce the weight of the device. Accelerators control system, which today is in the demo phase, will also be significantly upgraded. This paper describes the stages of the IntraLine accelerator development, its current state and plans for the future R&D work, within the scope of our new Intra-Dose project.