Mechanical Reliability Evaluation of an Oral Implant-Abutment System According to UNI EN ISO 14801 Fatigue Test Protocol
Authors:
- Enrico MarchettiEMEnrico Marchetti
- Stefano RattaSRStefano Ratta
- Stefano MummoloSMStefano Mummolo
- Simona TeccoSTSimona Tecco
- Raffaella PecciRaffaella Pecci
- Rossella BediniRBRossella Bedini
- Giuseppe MarzoGMGiuseppe Marzo
- 1 October 2016
- journal article
- Published by Wolters Kluwer Health in Implant Dentistry
- Vol. 25 (5) , 613-618
Abstract
The aim of this in vitro study was to evaluate the mechanical reliability of a dental implant system by testing its maximum fracture load and mechanical performance under cyclic fatigue stress. An experimental study according to the international standards (UNI EN ISO 14801: 2008) was performed using 13 implants (3.80 mm in diameter and 12 mm in length) with straight titanium abutments tightened to 30 N. Five samples were subjected to compression stress at break. Based on the mean fracture load value obtained in this test, the levels of dynamic loading range were set and were carried on at a frequency of 15 Hz for 5 × 10 cycles. The compression stress at break mean value of the tested implants was 430 N (SD ± 35.66 N). In the mechanical fatigue stress test, the fatigue limit for 5 × 10 load cycles was 172 N. The evaluated implant system proved to withstand considerable mechanical loads under the "worst-case" loading situation performed according to UNI EN ISO 14801 standard. The reliability of this test protocol makes it suitable to be accomplished for understanding and comparing mechanical properties of implant systems.Keywords
This publication has 13 references indexed in Scilit:
- Influence of the interface design on the yield force of the implant–abutment complex before and after cyclic mechanical loadingJournal of Prosthodontic Research, 2012
- Bending moments of zirconia and titanium abutments with internal and external implant–abutment connections after aging and chewing simulationClinical Oral Implants Research, 2011
- Implant–abutment interface design affects fatigue and fracture strength of implantsClinical Oral Implants Research, 2008
- In vitro differences of stress concentrations for internal and external hex implant–abutment connections: a short communicationJournal of Oral Rehabilitation, 2006
- Fatigue resistance of two implant/abutment joint designsThe Journal of Prosthetic Dentistry, 2002
- Micromotion and dynamic fatigue properties of the dental implant–abutment interfaceThe Journal of Prosthetic Dentistry, 2001
- Mechanical complications of dental implantsClinical Oral Implants Research, 2000
- A 5‐year prospective clinical study of Astra Tech dental implants supporting fixed bridges or overdentures in the edentulous mandibleClinical Oral Implants Research, 1997
- An in vitro evaluation of the strength of an internal conical interface compared to a butt joint interface in implant designClinical Oral Implants Research, 1997
- A 3-year follow-up study on single implant treatmentJournal of Dentistry, 1993