First implementation of an innovative infra‐red camera system integrated into a mobile CBCT scanner for applicator tracking in brachytherapy—Initial performance characterization
Authors:
Open Access
- 16 April 2024
- journal article
- research article
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 25 (7) , e14364
Abstract
Purpose: To enable a real‐time applicator guidance for brachytherapy, we used for the first time infra‐red tracking cameras (OptiTrack, USA) integrated into a mobile cone‐beam computed tomography (CBCT) scanner (medPhoton, Austria). We provide the first description of this prototype and its performance evaluation.Methods: We performed assessments of camera calibration and camera‐CBCT registration using a geometric calibration phantom. For this purpose, we first evaluated the effects of intrinsic parameters such as camera temperature or gantry rotations on the tracked marker positions. Afterward, calibrations with various settings (sample number, field of view coverage, calibration directions, calibration distances, and lighting conditions) were performed to identify the requirements for achieving maximum tracking accuracy based on an in‐house phantom. The corresponding effects on camera‐CBCT registration were determined as well by comparing tracked marker positions to the positions determined via CBCT. Long‐term stability was assessed by comparing tracking and a ground‐truth on a weekly basis for 6 weeks.Results: Robust tracking with positional drifts of 0.02 ± 0.01 mm was feasible using the system after a warm‐up period of 90 min. However, gantry rotations affected the tracking and led to inaccuracies of up to 0.70 mm. We identified that 4000 samples and full coverage were required to ensure a robust determination of marker positions and camera‐CBCT registration with geometric deviations of 0.18 ± 0.03 mm and 0.42 ± 0.07 mm, respectively. Long‐term stability showed deviations of more than two standard deviations from the initial calibration after 3 weeks.Conclusion: We implemented for the first time a standalone combined camera‐CBCT system for tracking in brachytherapy. The system showed high potential for establishing corresponding workflows.Keywords
This publication has 48 references indexed in Scilit:
- Cone‐beam CT imaging with laterally enlarged field of view based on independently movable source and detectorMedical Physics, 2023
- Modern Tools for Modern BrachytherapyClinical Oncology, 2023
- Quality assurance and long‐term stability of a novel 3‐in‐1 X‐ray system for brachytherapyJournal of Applied Clinical Medical Physics, 2022
- Toward 3D-TRUS image-guided interstitial brachytherapy for cervical cancerBrachytherapy, 2022
- Technical evaluation of the cone‐beam computed tomography imaging performance of a novel, mobile, gantry‐based X‐ray system for brachytherapyJournal of Applied Clinical Medical Physics, 2021
- Ultrasound-guided Brachytherapy for Cervix CancerClinical Oncology, 2021
- Computed tomography versus magnetic resonance imaging in high-dose-rate prostate brachytherapy planning: The impact on patient-reported health-related quality of lifeBrachytherapy, 2020
- Comparison of computed tomography and magnetic resonance imaging in cervical cancer brachytherapy: A systematic reviewBrachytherapy, 2017
- Nine‐degrees‐of‐freedom flexmap for a cone‐beam computed tomography imaging device with independently movable source and detectorMedical Physics, 2016
- Combining transrectal ultrasound and CT for image-guided adaptive brachytherapy of cervical cancer: Proof of conceptBrachytherapy, 2016