Evaluation of scattered radiation from fluoroscopy using small OSL dosimeters
Authors:
- 6 November 2019
- journal article
- research article
- Published by Springer Nature in Radiological Physics and Technology
- Vol. 12 (4) , 393-400
Abstract
Recently, there has been a significant amount of interest in studying the importance of radiation doses to the eye lens during endoscopic retrograde cholangiopancreatography (ERCP). A study that focused on measuring the scattered radiation using an ionization chamber survey meter reported that a lead curtain was useful in reducing the scattered radiation. The over-couch X-ray tube system tends to deliver higher doses to the head and neck of the staff involved in the procedure than the under-couch X-ray tube position. In this study, a small optically stimulated luminescence (OSL) dosimeter called the nanoDot was used to evaluate and measure the amount of radiation; this dosimeter was developed by Landauer Ltd. and was specifically designed for point measurements. There are numerous studies that have reported the usefulness of personal OSL dosimeters other than the nanoDot to measure scattered radiation. Here, we evaluated the amount of scattered radiation, along with the degree of reduction achieved with the use of a protective curtain, while employing a personal OSL dosimeter and nanoDot. When the scattered radiation dose was measured using the nanoDot, the maximum recorded value without a protective curtain was 0.363 mGy and that with a protective curtain was 0.026 mGy, both at the height of 100 cm. The maximum reduction rate of scattered radiation while using a protective curtain was approximately 93% and 97% at 100 cm and 150 cm, respectively. The measured values recorded using both personal OSL dosimeters and nanoDot machine were strongly correlated.Keywords
This publication has 9 references indexed in Scilit:
- Evaluation of Usefulness of X-ray Tube Protector for X-ray Fluoroscopy in ERCP etcJOURNAL OF THE JAPANESE ASSOCIATION OF RURAL MEDICINE, 2014
- Occupational radiation dose to eyes from endoscopic retrograde cholangiopancreatography procedures in light of the revised eye lens dose limit from the International Commission on Radiological ProtectionBritish Journal of Radiology, 2013
- ICRP PUBLICATION 118: ICRP Statement on Tissue Reactions and Early and Late Effects of Radiation in Normal Tissues and Organs — Threshold Doses for Tissue Reactions in a Radiation Protection ContextAnnals of the ICRP, 2012
- A head phantom study for intraocular dose evaluation of 64-slice multidetector CT examination in patients with suspected cranial traumaEuropean Journal of Radiology, 2011
- Angular dependence of the nanoDot OSL dosimeterMedical Physics, 2011
- Radiation Protection Basics for IVR StaffJapanese Journal of Radiological Technology, 2008
- Characterization of optically stimulated luminescent dosimeters, OSLDs, for clinical dosimetric measurementsMedical Physics, 2007
- Creation of a Crystalline Lens Radiation Exposure Defense Cover and the Effect of Radiation Exposure Decrease on Neuro-interventionsJapanese Journal of Radiological Technology, 2007
- Measurement of Scatter Radiation on MDCT Equipment Using an OSL DosimeterJapanese Journal of Radiological Technology, 2004