Experimental Evaluation of Relationship Between Radiofrequency Heating Near Implanted Conductive Devices, Scanner-Reported B1+rms, and Transmit Power
Open Access
- 11 April 2025
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Journal of Microwaves
- Vol. 5 (3) , 518-525
Abstract
Time-varying radiofrequency (RF) fields necessary to perform magnetic resonance imaging (MRI) may induce excessive heating near implanted conductive medical devices during MRI. The time and space-averaged root-mean-square effective magnetic field (B1+rms) and specific absorption rate (SAR) have been proposed as metrics to control the RF-induced heating and avoid unintended thermal injury. We experimentally evaluate the relationship between the RF-induced heating near an implanted conductive medical device, scanner-reported B1+rms, and RF power. RF heating was measured near the electrodes of a commercial deep brain stimulation (DBS) lead placed in a tissue-equivalent gel phantom using fluoroptic temperature probes in a commercial 3T scanner during MRI. Four RF transmit/receive coil combinations were used: a circularly polarized head transmit/receive coil, a 20-channel head/neck, a 32-channel head, or a 64-channel head/neck receive-only coil with a whole-body transmit coil. RF heating was induced by a 2D GRE sequence using two RF pulse types (fast and normal), three flip angles (30°, 60°, and 90°), and turning the receive-only coils off/on. The scanner-reported B1+rms and RF power were recorded. Measurements show that temperature change correlates linearly with both RF power and square of B1+rms for each coil and combination. However, the variation in heating for various RF coils and combinations was much larger for B1+rms compared to RF power. Additional studies across other MR scanners are needed to better understand the extent of variation in RF-induced heating near implanted conductive devices as a function of scanner-reported B1+rms and RF power to develop conservative and reliable patient labeling.Keywords
All Related Versions
Funding Information
- Zuckerman Mind Brain Behavior Institute
- Columbia University, New York (10027)
- Zuckerman Mind Brain Behavior Institute MRI Platform
- Columbia MR Research Center at Columbia University
- Rockefeller Neuroscience Institute
- West Virginia University
This publication has 24 references indexed in Scilit:
- Magnetic Resonance Imaging in Patients With Cardiac Implantable Electronic Devices With Abandoned LeadsJAMA Cardiology, 2021
- Temperatures in Pigs During 3 T MRI Temperatures, Heart Rates, and Breathing Rates of Pigs During RF Power Deposition in a 3 T (128 MHz) Body CoilBioelectromagnetics, 2020
- MRI‐Related Heating of Implants and Devices: A ReviewJournal of Magnetic Resonance Imaging, 2020
- Safety of magnetic resonance imaging in patients with legacy pacemakers and defibrillators and abandoned leadsHeart Rhythm, 2018
- In vivo radiofrequency heating in swine in a 3T (123.2-MHz) birdcage whole body coilMagnetic Resonance in Medicine, 2013
- Radiofrequency heating in porcine models with a “large” 32 cm internal diameter, 7 T (296 MHz) head coilMagnetic Resonance in Medicine, 2011
- Effect of the extracranial deep brain stimulation lead on radiofrequency heating at 9.4 Tesla (400.2 MHz)Journal of Magnetic Resonance Imaging, 2010
- Radiofrequency heating at 9.4T: In vivo temperature measurement results in swineMagnetic Resonance in Medicine, 2007
- Neurostimulation systems for deep brain stimulation: In vitro evaluation of magnetic resonance imaging–related heating at 1.5 teslaJournal of Magnetic Resonance Imaging, 2002
- Orthopedic implant devices: Prevalence and sociodemographic findings from the 1988 National Health Interview SurveyJournal of Applied Biomaterials, 1991