A determination of the Planck constant by the generalized joule balance method with a permanent-magnet system at NIM
- 13 January 2016
- journal article
- research article
- Published by IOP Publishing in Metrologia
- Vol. 53 (1) , 86-97
Abstract
The joule balance experiment has been carried out at the National Institute of Metrology, China (NIM) since 2007. By the end of 2013 the first generation of the joule balance (NIM-1) achieved a measurement uncertainty of 7.2 × 10−6 (k = 1). To reduce the measurement uncertainty further, the next generation of the joule balance apparatus (NIM-2) system is under construction. A new coil system using ferromagnetic material is being adopted in NIM-2 to reduce self-heating in the coils. However, the effects on the measurement of the mutual inductance from the nonlinearity and hysteresis of the ferromagnetic material will bring a considerable measurement uncertainty. Inspired by the watt balance, the measurement of the mutual inductance is replaced by an equivalent measurement of the magnetic flux linkage difference. The nonlinearity and hysteresis will not be a problem in the measurement of the magnetic flux linkage difference. This technique comes from the watt balance method. It is called the generalized joule balance method, which is actually a modification of the watt balance method. However, it still represents a valid change that can reduce the difficulty of dynamic measurement experienced using the watt balance. Permanent magnets can also be adopted in the generalized joule balance. To check the feasibility of the generalized joule balance method, some preliminary experiments have been performed on NIM-1. A yokeless permanent magnet system has been designed and used to replace the exciting coils in NIM-1. In this paper, the structure of the yokeless permanent magnet system is introduced. Furthermore, a determination of the Planck constant with the permanent magnet system is presented. The value of the Planck constant h we obtained is 6.626 069(17) × 10−34 J s with a relative standard uncertainty of 2.6 × 10−6.Keywords
Funding Information
- The key support program of major research project of China National Natural Science Foundation (91536224)
- The China National Key Scientific Instrument and Equipment Development Project (2011YQ090004)
This publication has 26 references indexed in Scilit:
- Coils and the Electromagnet Used in the Joule Balance at the NIMIEEE Transactions on Instrumentation and Measurement, 2015
- A nonlinearity in permanent-magnet systems used in watt balancesMetrologia, 2014
- High-accuracy measurement of linearity of optical detectors based on flux addition of LEDs in an integrating sphereMetrologia, 2013
- Nonlinear magnetic error evaluation of a two-mode watt balance experimentMetrologia, 2013
- Design of the Permanent-Magnet System for NIST-4IEEE Transactions on Instrumentation and Measurement, 2012
- Mutual inductance measurement of the superconducting coil for Joule Balance2012
- Precisely measuring the Planck constant by electromechanical balancesMeasurement, 2012
- Recent Development on the Joule Balance at NIMIEEE Transactions on Instrumentation and Measurement, 2011
- Towards a better definition of the kilogramMetrologia, 2011
- Progress on the BIPM Watt BalanceIEEE Transactions on Instrumentation and Measurement, 2008