SUP-Net: Slow-Time Upsampling Network for Aliasing Removal in Doppler Ultrasound
Open Access
- 24 July 2025
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Medical Imaging
- Vol. 45 (1) , 83-97
Abstract
Doppler ultrasound modalities, which include spectral Doppler and color flow imaging, are frequently used tools for flow diagnostics because of their real-time point-of-care applicability and high temporal resolution. When implemented using pulse-echo sensing and phase shift estimation principles, this modality’s pulse repetition frequency (PRF) is known to influence the maximum detectable velocity. If the PRF is inevitably set below the Nyquist limit due to imaging requirements or hardware constraints, aliasing errors or spectral overlap may corrupt the estimated flow data. To solve this issue, we have devised a deep learning-based framework, powered by a custom slow-time upsampling network (SUP-Net) that leverages spatiotemporal characteristics to upsample the received ultrasound signals across pulse echoes acquired using high-frame-rate ultrasound (HiFRUS). Our framework infers high-PRF signals from signals acquired at low PRF, thereby improving Doppler ultrasound’s flow estimation quality. SUP-Net was trained and evaluated on in vivo femoral acquisitions from 20 participants and was applied recursively to resolve scenarios with excessive aliasing across a range of PRFs. We report the successful reconstruction of slow-time signals with frequency content that exceeds the Nyquist limit once and twice. By operating on the fundamental slow-time signals, our framework can resolve aliasing-related artifacts in several downstream modalities, including color Doppler and pulse wave Doppler.Keywords
Funding Information
- Natural Sciences and Engineering Council of Canada (CREATE-528202-2019)
- Natural Sciences and Engineering Council of Canada (RGPIN-2022-04157)
- Canadian Institutes of Health Research (PJT-153240)
- E.W.R. Steacie Memorial Fellowship (SMFSU-556263-2021)
- Canadian Space Agency (21FAWATA08)
This publication has 42 references indexed in Scilit:
- Optical coherence tomography angiographyProgress in Retinal and Eye Research, 2018
- Advanced flow MRI: emerging techniques and applicationsClinical Radiology, 2016
- U-Net: Convolutional Networks for Biomedical Image Segmentation2015
- Ultrasound Medical Imaging2014
- Unsupervised dealiasing and denoising of color-Doppler dataMedical Image Analysis, 2011
- Point-of-Care UltrasonographyNew England Journal of Medicine, 2011
- Cooperation of fuzzy segmentation operators for correction aliasing phenomenon in 3D color Doppler imagingArtificial Intelligence in Medicine, 2000
- Computing requirements of modern medical diagnostic ultrasound machinesParallel Computing, 1998
- Faster femoral artery blood velocity kinetics at the onset of exercise following short-term trainingCardiovascular Research, 1996
- Clinical evaluation of a new anti-aliasing technique for ultrasound pulsed Doppler analysisUltrasound in Medicine & Biology, 1989