Region of interest selection and volume averaging for Gamma Knife radiochromic film dosimetry
Open Access
- 28 November 2025
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 70 (23) , 235024
Abstract
Objective. Accurate determination of relative output factors (ROFs) is paramount in stereotactic radiosurgery dosimetry, particularly for small fields employed in Leksell Gamma Knife (LGK) systems. Existing guidelines for LGK film-based dosimetry lack specificity regarding the selection of a region of interest (ROI) for ROF measurements. Approach. This investigation assessed the dosimetric performance of Gafchromic TM EBT4 film, focusing on establishing a quantitative metric for the unambiguous selection of a small ROI, defined relative to the beam’s full width at half maximum (FWHM), for LGK ROF measurements. Film irradiations were performed on an LGK ICON unit using 16-, 8-, and 4 mm collimators. Calibration curves were generated by irradiating 15 films with 16 mm beams in 15 s increments, with films subsequently scanned in both portrait and landscape orientations at various resolutions. Dosimetric accuracy was assessed using the net optical density transfer function along with corresponding uncertainty propagation. The influence of detector size on ROF determination was calculated using ROIs with diameters systematically varied to include 1/10, 1/5, 1/4, 1/3, 2/5, 1/2, 2/3, and 3/4 of the beam’s FWHM. Main results. ROF measurements demonstrated high sensitivity to ROI size. A metric employing a ¼ FWHM ROI yielded ROFs values within 1% of the established Leksell GammaPlan (LGP) reference data, effectively minimizing film volume-averaging effects. Systematic dose errors, up to ±20% on average, were observed due to orientation mismatches between calibration and measurement scans. Furthermore, across all evaluated scan resolutions both FWHM and penumbra measurements consistently conformed to LGP specifications, specifically within ±0.5 mm and ±1 mm, respectively. Significance. The ¼ FWHM ROI provides a clinically practical method for LGK ROF determination, addressing the absence of a clear ROI definition in TG-178 report. This metric achieves a suitable balance for film-based measurements, ensuring a sufficient number of ROI pixels for statistical significance while minimizing volume-averaging effects.Keywords
This publication has 28 references indexed in Scilit:
- Effects of rotation and scanning spatial resolution on gafchromic film response in high energy photon beamApplied Radiation and Isotopes, 2025
- Characterization of scanning orientation and lateral response artifact for EBT4 Gafchromic filmJournal of Applied Clinical Medical Physics, 2023
- Monte-Carlo techniques for radiotherapy applications I: introduction and overview of the different Monte-Carlo codesJournal of Radiotherapy in Practice, 2023
- Report of AAPM Task Group 155: Megavoltage photon beam dosimetry in small fields and non‐equilibrium conditionsMedical Physics, 2021
- Recommendations on the practice of calibration, dosimetry, and quality assurance for gamma stereotactic radiosurgery: Report of AAPM Task Group 178Medical Physics, 2021
- Dosimetry of small static fields used in external photon beam radiotherapy: Summary of TRS‐483, the IAEA–AAPM international Code of Practice for reference and relative dose determinationMedical Physics, 2018
- Reference radiochromic film dosimetry: Review of technical aspectsPhysica Medica, 2016
- Monte Carlo calculated and experimentally determined output correction factors for small field detectors in Leksell Gamma Knife Perfexion beamsPhysics in Medicine & Biology, 2015
- FLUKA Monte Carlo simulation for the Leksell Gamma Knife Perfexion radiosurgery system: Homogeneous mediaPhysica Medica, 2013
- Monte Carlo simulation of the Leksell Gamma Knife$reg$: I. Source modelling and calculations in homogeneous mediaPhysics in Medicine & Biology, 2002