Investigation and compensation of the nonlinear response in photomultiplier tubes for quantitative single-shot measurements
- 1 March 2012
- journal article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 83 (3) , 034901
Abstract
A concept for time-sensitive optical detectors is described that shows how to confirm whether the detection device is operating in the linear response regime. By evaluating the recorded time decay of a thermographic phosphor, even weak saturation effects far from obvious situations can be identified and further related to either optical or electrical saturation. The concept has been validated by running a PMT detector close to saturation and exposing it to the optical signal decay of two different thermographic phosphors, La2O2S:Eu and CdWO4. It was confirmed that short but intense light exposures at the beginning of an individual time decay influence the detector response for the rest of the decaying signal including temporal areas, where the anode current has dropped well below the manufacturer specified current limit. Such situations are common when applying, e.g., phosphor thermometry where it is necessary to retrieve the full decay curve from a single-shot event, i.e., standard techniques based on single-photon counting are omitted. Finally, means of compensation are introduced in order to facilitate the retrieval of useful information from the measurement data when operation in the non-linear response regime is inevitable.Keywords
This publication has 12 references indexed in Scilit:
- Limitations of ICCD detectors and optimized 2D phosphor thermometryMeasurement Science and Technology, 2012
- Influence of spatial laser energy distribution on evaluated soot particle sizes using two-colour laser-induced incandescence in a flat premixed ethylene/air flameApplied Physics B, 2009
- An algorithm for the characterisation of multi-exponential decay curvesOptics and Lasers in Engineering, 2009
- Investigations of blue emitting phosphors for thermometryMeasurement Science and Technology, 2008
- 2D-temperature imaging of single droplets and sprays using thermographic phosphorsApplied Physics B, 2004
- Fast pulsed UV light source and calibration of non-linear photomultiplier responseNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2003
- Pulse mode saturation properties of photomultiplier tubesReview of Scientific Instruments, 1978
- Methods for drift stabilization and photomultiplier linearization for photometric ellipsometers and polarimetersReview of Scientific Instruments, 1978
- Linearitätsabweichungen bei Strahlungsmessungen mit PhotovervielfachernApplied Optics, 1972
- Representation of the Temperature Scale above 1337.58 K with Photoelectric Direct Current PyrometersMetrologia, 1969