Decreased Delivery of Inappropriate Shocks Achieved by Remote Monitoring of ICD: A Substudy of the ECOST Trial
- 10 April 2014
- journal article
- research article
- Published by Wiley in Journal of Cardiovascular Electrophysiology
- Vol. 25 (7) , 763-770
Abstract
Inappropriate Shocks Reduction by Remote ICD Monitoring: Introduction: Inappropriate shocks remain a highly challenging complication of implantable cardioverter defibrillators (ICD). We examined whether automatic wireless remote monitoring (RM) of ICD, by providing early notifications of triggering events, lowers the incidence of inappropriate shocks.Methods and results: We studied 433 patients randomly assigned to RM (n = 221; active group) versus ambulatory follow‐up (n = 212; control group). Patients in the active group were seen in the ambulatory department once a year, unless RM reported an event requiring an earlier ambulatory visit. Patients in the control group were seen in the ambulatory department every 6 months. The occurrence of first and further inappropriate shocks, and their causes in each group were compared.The characteristics of the study groups, including pharmaceutical regimens, were similar. Over a follow‐up of 27 months, 5.0% of patients in the active group received ≥1 inappropriate shocks versus 10.4% in the control group (P = 0.03). A total of 28 inappropriate shocks were delivered in the active versus 283 in the control group. Shocks were triggered by supraventricular tachyarrhythmias (SVTA) in 48.5%, noise oversensing in 21.2%, T wave oversensing in 15.2%, and lead dysfunction in 15.2% of patients. The numbers of inappropriate shocks delivered per patient, triggered by SVTA and by lead dysfunction, were 74% and 98% lower, respectively, in the active than in the control group.Conclusion: RM was highly effective in the long‐term prevention of inappropriate ICD shocks.Keywords
Funding Information
- Biotronik SE & Co. KG.
This publication has 14 references indexed in Scilit:
- Reduction in Inappropriate Therapy and Mortality through ICD ProgrammingNew England Journal of Medicine, 2012
- A randomized study of remote follow-up of implantable cardioverter defibrillators: safety and efficacy report of the ECOST trialEuropean Heart Journal, 2012
- Comparison Between Atrial Fibrillation‐Triggered Implantable Cardioverter‐Defibrillator (ICD) Shocks and Inappropriate Shocks Caused by Lead Failure: Different Impact on Prognosis in Clinical PracticeJournal of Cardiovascular Electrophysiology, 2012
- Inappropriate Implantable Cardioverter-Defibrillator ShocksJACC, 2011
- Efficacy and Safety of Automatic Remote Monitoring for Implantable Cardioverter-Defibrillator Follow-UpCirculation, 2010
- Contributions of remote monitoring to the follow-up of implantable cardioverter-defibrillator leads under advisoryEuropean Heart Journal, 2010
- ACC/AHA/HRS 2008 Guidelines for Device-Based Therapy of Cardiac Rhythm Abnormalities: Executive SummaryCirculation, 2008
- Inappropriate Implantable Cardioverter-Defibrillator Shocks in MADIT IIJACC, 2008
- Remote, Wireless, Ambulatory Monitoring of Implantable Pacemakers, Cardioverter Defibrillators, and Cardiac Resynchronization Therapy Systems: Analysis of a Worldwide DatabasePacing and Clinical Electrophysiology, 2007
- Frequency and Causes of Implantable Cardioverter-Defibrillator Therapies: Is Device Therapy Proarrhythmic?The American Journal of Cardiology, 2006