Progressive-toric IOL design reduces residual astigmatism with increasing pupil size: a ray-tracing simulation based on corneal topography data
- 23 February 2021
- journal article
- research article
- Published by Optica Publishing Group in Biomedical Optics Express
- Vol. 12 (3) , 1568-1576
Abstract
Population studies indicate that astigmatism decreases from the corneal center toward the periphery. A standard toric intraocular lens (IOL) with a constant cylinder power cannot correct uniformly across this gradient. We built an astigmatic eye model based on corneal topography data. A progressive-toric lens with gradually decreasing cylinder power was compared with an identically designed lens but featuring conventional astigmatism correction. Residual astigmatism did not differ significantly (P=0.06) at 3 mm, and the Strehl ratio was identical for both lenses (0.51 ±0.15, P=0.88). At 5 mm, the progressive IOL yielded significantly lower residual astigmatism by 0.10 D (P<0.001). The Strehl ratio was 0.30 ±0.08 with the progressive and 0.29 ±0.08 with the standard lens (P<0.001). At 3 mm, the optical performance was comparable for both IOLs. However, at 5 mm, the progressive-toric was more effective in correcting astigmatism, and it yielded reduced residual astigmatism compared to a standard toric lens.Keywords
Funding Information
- Klaus Tschira Stiftung
- Kowa Pharmaceutical Europe
This publication has 37 references indexed in Scilit:
- Astigmatism correction with toric intraocular lenses: wavefront aberrometry and quality of lifeBritish Journal of Ophthalmology, 2013
- Contribution of posterior corneal astigmatism to total corneal astigmatismJournal of Cataract & Refractive Surgery, 2012
- Implantation of a multifocal toric intraocular lens with a surface-embedded near segment after repeated LASIK treatmentsJournal of Cataract & Refractive Surgery, 2012
- Residual Astigmatism Following Toric Intraocular Lens Implantation Related to Pupil SizeJournal of Refractive Surgery, 2012
- Impact of axis misalignment of toric intraocular lenses on refractive outcomes after cataract surgeryJournal of Cataract & Refractive Surgery, 2010
- Comparison of toric intraocular lenses and peripheral corneal relaxing incisions to treat astigmatism during cataract surgeryJournal of Cataract & Refractive Surgery, 2010
- Tilt and decentration of spherical and aspheric intraocular lenses: Effect on higher-order aberrationsJournal of Cataract & Refractive Surgery, 2009
- The Topography of the Central and Peripheral CorneaInvestigative Ophthalmology & Visual Science, 2006
- Intraindividual comparison of higher-order aberrations after implantation of aspherical and spherical intraocular lenses as a function of pupil diameterJournal of Cataract & Refractive Surgery, 2006
- Metrics of optical quality derived from wave aberrations predict visual performanceJournal of Vision, 2004