Dosimetric accuracy of planning and delivering small proton therapy fields
Authors:
- Juliane DaartzJDJuliane Daartz
- Harald PaganettiHPHarald Paganetti
- 19 November 2010
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 55 (24) , 7425-7438
Abstract
A detailed Monte Carlo model of a proton therapy treatment head was commissioned to simulate the delivery of small field proton treatments. Small fields are challenging in the planning and quality assurance process because of aperture scattering and dosimetric disequilibrium. Four patients with small fields used in all or parts of their treatment course were studied, including two stereotactic patients and two fractionated patients. For the two stereotactic patients the overall difference of the dose covering at least 95% of the gross tumor volume between the Monte Carlo calculations and the delivery was -0.2% and -1.6%, respectively. For the two fractionated patients the overall difference of the dose covering at least 95% of the clinical target volume was -3.0% and 1.0%, respectively. We have thus confirmed that our current planning and delivery process for small proton fields is accurate enough to treat small lesions in the patient. Furthermore, we studied the impact of field size corrections and identified limitations of the pencil beam algorithm for predicting hot and cold spots and range degradation in the target due scattering in heterogeneities. For the four cases studied in this paper, these limitations appear to impact individual field calculations, but do not have a significant impact on the prescribed dose over multiple fields.Keywords
This publication has 16 references indexed in Scilit:
- Dose to water versus dose to medium in proton beam therapyPhysics in Medicine & Biology, 2009
- Field size dependence of the output factor in passively scattered proton therapy: Influence of range, modulation, air gap, and machine settingsMedical Physics, 2009
- Commissioning a passive‐scattering proton therapy nozzle for accurate SOBP deliveryMedical Physics, 2009
- Clinical implementation of full Monte Carlo dose calculation in proton beam therapyPhysics in Medicine & Biology, 2008
- Monte Carlo investigation of collimator scatter of proton-therapy beams produced using the passive scattering methodPhysics in Medicine & Biology, 2007
- The prediction of output factors for spread-out proton Bragg peak fields in clinical practicePhysics in Medicine & Biology, 2005
- Adaptation of GEANT4 to Monte Carlo dose calculations based on CT dataMedical Physics, 2004
- Accurate Monte Carlo simulations for nozzle design, commissioning and quality assurance for a proton radiation therapy facilityMedical Physics, 2004
- A pencil beam algorithm for proton dose calculationsPhysics in Medicine & Biology, 1996
- Evaluation of a pencil-beam dose calculation technique for charged particle radiotherapyInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 1996