Objective: Occlusive cuff inflation in ambulatory blood pressure (BP) monitoring disturbs the daily life of the user, and affects efficacy of monitoring. To overcome this limitation, we have developed a novel minimally-occlusive cuff method for stress-free measurement of BP. This study aimed to experimentally evaluate the reliability of this method. Methods In this method, a thin-plate-type ultrasound probe placed between the cuff and the skin is used to measure the ultrasonic dimension of the artery. Analyzing the arterial dimension and the cuff pressure measured during mild cuff inflation (cuff pressure $\lt50$ mmHg) allows estimation of systolic ( $\rm SBP_e$ ) and diastolic BP ( $\rm DBP_e$ ). We evaluated this method in 6 anesthetized dogs by attaching the cuff with the probe to the right thigh to get $\rm SBP_e$ and $\rm DBP_e$ . We measured reference SBP and DBP using an intra-arterial catheter and determined the pulse arrival time (PAT) commonly employed in cuff-less BP monitoring. BP was perturbed by infusing noradrenaline or nitroprusside. Results: $\rm DBP_e$ correlated tightly with DBP with a coefficient of determination ( $\rm R^2$ ) of 0.85 ± 0.08, and predicted DBP with a bias of 3.9 ± 7.9 mmHg after a one-time calibration. Reciprocal of PAT correlated poorly with DBP ( $\rm R^2$ = 0.49 ± 0.17), and predicted DBP less accurately than this method. $\rm SBP_e$ correlated well with SBP ( $\rm R^2$ = 0.78 ± 0.08) Conclusion: This method reliably tracks BP changes without occlusive cuff inflation. Once calibrated, this method measures DBP accurately. Significance: This method has the potential for stress-free BP measurement in ambulatory BP monitoring.
Funding Information
  • JSPS KAKENHI (15K01307, 18K12126)

This publication has 30 references indexed in Scilit: