Effect of Hollow Fiber Length on Solute Removal and Quantification of Internal Filtration Rate by Doppler Ultrasound
Authors:
- 1 February 2003
- journal article
- research article
- Published by SAGE Publications in The International Journal of Artificial Organs
- Vol. 26 (2) , 129-134
Abstract
Renal replacement therapy with dialyzers capable of enhanced internal filtration (IF) can be an alternative to standard hemodiafiltration, as it provides convective solute removal comparable to that of hemodiafiltration by a simple procedure. In this study, we clinically evaluated the effect of the hollow fiber length in the dialyzer, a crucial factor influencing the rate of IF, by comparing two commercial dialyzers (BS-1.6U, BS-1.6UL, Toray, Japan) which differed in the fiber length, but had the same surface area and inner diameter of their hollow fibers. We showed that in the dialyzer with the longer fibers, the pressure profile along the dialyzer was significantly altered, and the solute clearance tended to be increased. In addition, we successfully quantified the IF rate with a Doppler ultrasound in the experimental circuit, by measuring the blood flow velocities along the bundle of fibers. We showed that the changes in the blood flow velocity were more marked in the dialyzer with the longer fibers; the calculated IF rates in the dialyzers with the shorter and longer fibers were 11.1 mL/min and 37.7 mL/min, respectively, which seemed to be compatible with the solute clearances. This simple and readily applicable method is expected to be useful in the development of modified dialyzers to fully exploit the benefits of IF in renal replacement therapy.Keywords
This publication has 13 references indexed in Scilit:
- Effects of Hematocrit and Blood Flow Distribution on Solute Clearance in Hollow-Fiber HemodialyzersNephron, 2001
- Effects of a reduced inner diameter of hollow fibers in hemodialyzersKidney International, 2000
- Effects of Internal Filtration on the Solute Removal Efficiency of a DialyzerASAIO Journal, 2000
- Comparison of mortality in ESRD patients on convective and diffusive extracorporeal treatmentsKidney International, 1999
- On-line haemodiafiltration: state of the artNephrology Dialysis Transplantation, 1998
- Internal filtration--advantage in haemodialysis?Nephrology Dialysis Transplantation, 1996
- Diffusive and convective transfer of cytokine-inducing bacterial products across hemodialysis membranesKidney International, 1995
- A new scintigraphic method to characterize ultrafiltration in hollow fiber dialyzersKidney International, 1992
- An In Vivo Analysis of Reverse Ultrafiltration During High-Flux and High-Efficiency DialysisAmerican Journal of Kidney Diseases, 1992
- Backfiltration in Hemodialyzers with Highly Permeable MembranesBlood Purification, 1984