Transverse oscillations for phased array vector velocity imaging
- 1 October 2010
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- No. 10510117,p. 1323-1327
Abstract
Medical ultrasound imaging is widely used to visualize blood flow in the human circulatory system. However, conventional methods are angle dependent. The Transverse Oscillation (TO) method is able to measure the lateral velocity component, and it has been demonstrated in in vivo measurements of superficial blood vessels. To broaden the usability of the method, it should be expanded to a phased array geometry enabling vector velocity imaging of the heart. Therefore, the scan depth has to be increased to 10-15 cm. This paper presents suitable pulse echo fields (PEΓ). Two lines are beamformed in receive to obtain lateral spatial in-phase and quadrature components. The relative mean bias and standard deviation of the lateral velocity component are computed as performance measures. For the PEF, the coefficient of variance, CV, of the spectral frequencies, and the energy ratio, ER, of leakage into negative frequencies are used as metrics to assess estimator performance. At 10 em's depth for an initial setup, the relative mean bias and standard deviation are 9.1% and 9.5%, respectively. At a depth of 15 cm, the values are 20% and 13%, respectively. The PEF metric ER can be used to assess the bias (correlation coefficient, R: -0.76), and therefore predict estimator performance. CV is correlated with the standard deviation (R=0.74). The results demonstrate the potential for using a phased array for vector velocity imaging at larger depths, and potentially for imaging the heart.Keywords
This publication has 8 references indexed in Scilit:
- Performance of the Transverse Oscillation method using beamformed data from a commercial scanner2009
- In vivo validation of a blood vector velocity estimator with MR angiographyIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 2009
- Investigation of transverse oscillation methodIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 2006
- Carotid Artery Stenosis: Gray-Scale and Doppler US Diagnosis—Society of Radiologists in Ultrasound Consensus ConferenceRadiology, 2003
- A new estimator for vector velocity estimation [medical ultrasonics]IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 2001
- A new method for estimation of velocity vectorsIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 1998
- Multi-dimensional velocity estimation with ultrasound using spatial quadratureIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 1998
- Calculation of pressure fields from arbitrarily shaped, apodized, and excited ultrasound transducersIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 1992