Estimation of intra-operator variability in perfusion parameter measurements using DCE-US
Authors:
- Marianne GauthierMGMarianne Gauthier
- Ingrid LeguerneyILIngrid Leguerney
- Jessie ThalmensiJTJessie Thalmensi
- Mohamed ChebilMCMohamed Chebil
- Sarah ParisotSPSarah Parisot
- Pierre PeronneauPPPierre Peronneau
- Alain RocheARAlain Roche
- Nathalie LassauNLNathalie Lassau
Open Access
- 1 January 2011
- journal article
- Published by Baishideng Publishing Group Inc. in World Journal of Radiology
- Vol. 3 (3) , 70-81
Abstract
To investigate intra-operator variability of semi-quantitative perfusion parameters using dynamic contrast-enhanced ultrasonography (DCE-US), following bolus injections of SonoVue(®). The in vitro experiments were conducted using three in-house sets up based on pumping a fluid through a phantom placed in a water tank. In the in vivo experiments, B16F10 melanoma cells were xenografted to five nude mice. Both in vitro and in vivo, images were acquired following bolus injections of the ultrasound contrast agent SonoVue(®) (Bracco, Milan, Italy) and using a Toshiba Aplio(®) ultrasound scanner connected to a 2.9-5.8 MHz linear transducer (PZT, PLT 604AT probe) (Toshiba, Japan) allowing harmonic imaging ("Vascular Recognition Imaging") involving linear raw data. A mathematical model based on the dye-dilution theory was developed by the Gustave Roussy Institute, Villejuif, France and used to evaluate seven perfusion parameters from time-intensity curves. Intra-operator variability analyses were based on determining perfusion parameter coefficients of variation (CV). In vitro, different volumes of SonoVue(®) were tested with the three phantoms: intra-operator variability was found to range from 2.33% to 23.72%. In vivo, experiments were performed on tumor tissues and perfusion parameters exhibited values ranging from 1.48% to 29.97%. In addition, the area under the curve (AUC) and the area under the wash-out (AUWO) were two of the parameters of great interest since throughout in vitro and in vivo experiments their variability was lower than 15.79%. AUC and AUWO appear to be the most reliable parameters for assessing tumor perfusion using DCE-US as they exhibited the lowest CV values.Keywords
This publication has 43 references indexed in Scilit:
- Reproducibility of Perfusion Parameters in Dynamic Contrast-Enhanced MRI of Lung and Liver Tumors: Effect on Estimates of Patient Sample Size in Clinical Trials and on Individual Patient ResponsesAmerican Journal of Roentgenology, 2010
- Early Quantitative Evaluation of a Tumor Vasculature Disruptive Agent AVE8062 Using Dynamic Contrast-Enhanced UltrasonographyInvestigative Radiology, 2008
- Experimental investigation and theoretical modelling of the nonlinear acoustical behaviour of a liver tissue and comparison with a tissue mimicking hydrogelJournal of Materials Science: Materials in Medicine, 2007
- Comparison of different mathematical models to analyze diminution kinetics of ultrasound contrast enhancement in a flow phantomUltrasound in Medicine & Biology, 2005
- Characterization of Focal Liver Lesions with Contrast-specific US Modes and a Sulfur Hexafluoride–filled Microbubble Contrast Agent: Diagnostic Performance and ConfidenceRadiology, 2004
- Direct evidence that the VEGF-specific antibody bevacizumab has antivascular effects in human rectal cancerNature Medicine, 2004
- High-resolution functional imaging with ultrasound contrast agents based on RF processing in an in vivo kidney experimentUltrasound in Medicine & Biology, 2001
- Characteristics of SonoVue™Echocardiography, 1999
- Dynamic CT measurement of cerebral blood flow: a validation study.1999
- Differential sensitivity of breast cancer and melanoma cells to proteasome inhibitor VelcadeInternational Journal of Molecular Medicine, 1998