Automatic Column Coupling System To Operate Chromatographic Supports Closer To Their Kinetic Performance Limit and To Enhance Method Development
- 30 December 2009
- journal article
- research article
- Published by American Chemical Society (ACS) in Analytical Chemistry
- Vol. 82 (3) , 1054-1065
Abstract
A new hardware solution is proposed that allows one to automatically change the length of a chromatographic bed. The setup is based on the serial coupling of chromatographic columns using two rotor−stator valves (with N positions, N + 1 ports). Despite the use of a prototype setup requiring rather long connection tubing, only 9% loss in efficiency is observed for compounds with a retention factor above 4 compared to the efficiency expected on the basis of the individual column results. It has been demonstrated for a number of isocratic and gradient separations that the system allows one to realize considerable analysis time savings by adapting the total column length to the specific sample requirements and/or to the stage of method development wherein one is working. During method development, a separation on a short column length can first be used to rapidly gain insight into the composition of the sample, leaving fewer runs to be done on a column of maximal length (offering efficiencies that are inaccessible with individual column systems). The ease with which information can be obtained on columns of different lengths can furthermore be exploited for screening purposes to detect coeluting components in a stage wherein they still appear completely unresolved (i.e., have a resolution well below Rs = 0.5).Keywords
This publication has 25 references indexed in Scilit:
- Investigation of the validity of the kinetic plot method to predict the performance of coupled column systems operated at very high pressures under different thermal conditionsJournal of Chromatography A, 2009
- Kinetic plot method as a tool to design coupled column systems producing 100,000 theoretical plates in the shortest possible timeJournal of Chromatography A, 2008
- Evaluation of the properties of a superficially porous silica stationary phase in hydrophilic interaction chromatographyJournal of Chromatography A, 2008
- Fused‐core particle technology as an alternative to sub‐2‐μm particles to achieve high separation efficiency with low backpressureJournal of Separation Science, 2007
- High efficiency liquid chromatography on conventional columns and instrumentation by using temperature as a variable: Kinetic plots and experimental verificationJournal of Chromatography A, 2007
- High-efficiency liquid chromatography on conventional columns and instrumentation by using temperature as a variable: I. Experiments with 25cm×4.6mm I.D., 5μm ODS columnsJournal of Chromatography A, 2006
- Method development in reversed-phase chromatography2003
- High-performance liquid chromatography of basic compoundsJournal of Chromatography A, 1998
- Practical HPLC Method Development1997
- Mode of operation and performance characteristics of microbore columns for use in liquid chromatographyJournal of Chromatography A, 1979