Modeling and characterization of AlN-Based piezoelectric micromachined ultrasonic transducers
- 22 November 2024
- journal article
- research article
- Published by IOP Publishing in Journal of Micromechanics and Microengineering
- Vol. 34 (12) , 125003
Abstract
In this paper, we present the modeling and characterization of AlN-based piezoelectric micromachined ultrasonic transducers (PMUTs) dedicated to biomedical applications. Our study encompasses detailed analytical modeling of these transducers in both air and water environments. The model is confronted with finite element simulations that account for realistic issues such as imperfect clamping. To validate the model’s relevance, we compare its predictions with results from optical, electrical, and acoustical characterization. This comprehensive approach provides a robust framework for understanding PMUT behavior and facilitates efficient modeling of future devices.Keywords
Funding Information
- Electronic Components and Systems for European Leadership
This publication has 20 references indexed in Scilit:
- Review of pMUTs for medical imaging: towards high frequency arraysBiomedical Physics & Engineering Express, 2023
- An Improved Lumped Element Model for Circular-Shape pMUTsIEEE Open Journal of Ultrasonics, Ferroelectrics, and Frequency Control, 2022
- Wafer-level experimental study of residual stress in AlN-based bimorph piezoelectric micromachined ultrasonic transducerEngineering Research Express, 2020
- Characterization of polymer-based piezoelectric micromachined ultrasound transducers for short-range gesture recognition applicationsJournal of Micromechanics and Microengineering, 2019
- Review of piezoelectric micromachined ultrasonic transducers and their applicationsJournal of Micromechanics and Microengineering, 2017
- System level modeling and design maps of PMUTs with residual stressesSensors and Actuators A: Physical, 2017
- Modeling and process design optimization of a piezoelectric micromachined ultrasonic transducers (PMUT) using lumped elements parametersMicrosystem Technologies, 2016
- Bimorph Piezoelectric Micromachined Ultrasonic TransducersJournal of Microelectromechanical Systems, 2016
- A top-crossover-to-bottom addressed segmented annular array using piezoelectric micromachined ultrasonic transducersJournal of Micromechanics and Microengineering, 2015
- Equivalent circuit-based analysis of CMUT cell dynamics in arraysIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 2013