Hydrogel hybrid porcine pericardium for the fabrication of a pre-mounted TAVI valve with improved biocompatibility
- 23 January 2019
- journal article
- research article
- Published by Royal Society of Chemistry (RSC) in Journal of Materials Chemistry B
- Vol. 7 (9) , 1427-1434
Abstract
Transcatheter aortic valve implantation (TAVI) has been developed years ago for patients who cannot undergo a surgical aortic valve replacement (SAVR). Although TAVI possesses the advantages of lower trauma and simpler manipulation compared to SAVR, the need for storage in glutaraldehyde (GLU) and a tedious intraoperative assembly process have caused great inconvenience for its further application. A pre-mounted TAVI valve assembled by mounting a dry valve frame to a delivery system is expected to address these problems. However, the currently used GLU treated leaflet cannot unfold normally after being crimped for a long-term and loses its function when the BHV is assembled to the catheter. Besides, its cytotoxicity and immune response after implantation are still problems to be solved. In the present study, a hydrogel hybrid porcine pericardium (HHPP) approach was developed to endow the BHVs with a favorable unfolding property and good biocompatibility. Three monomers with different charge characteristics (sodium acrylate, 2-methacryloyloxyethyl phosphorylcholine, and acryloyloxyethyltrimethyl ammonium chloride) were complexed with GLU treated PP (GLU-PP) to form three kinds of HHPPs (SAAH-PP, MPCH-PP, and DACH-PP). The results of the crimping simulation experiment showed that all HHPPs could quickly recover in PBS after being folded for 10 days, while the traditional BHVs (GLU-PP) could not recover under the same conditions. Bovine serum albumin adsorption and platelet adhesion test showed that SAAH-PP and MPCH-PP had good anti-adhesion abilities. A cell culture study indicated that all the three HHPPs promoted HUVEC growth and proliferation. In vivo biocompatibility studies showed that the immune response induced by MPCH-PP was reduced compared to that by GLU-PP. These studies demonstrated that the strategy of MPC hydrogel hybridization may be an effective approach to prepare a pre-mounted TAVI valve with improved biocompatibility.Keywords
Funding Information
- National Natural Science Foundation of China (31700833)
- China Association for Science and Technology (2017QNRC001)
This publication has 41 references indexed in Scilit:
- Reverse atom transfer radical polymerization (RATRP) for anti-clotting PU-LaCl3-g-P(MPC) filmsApplied Surface Science, 2012
- Porcine vena cava as an alternative to bovine pericardium in bioprosthetic percutaneous heart valvesBiomaterials, 2012
- Immune responses to implants – A review of the implications for the design of immunomodulatory biomaterialsBiomaterials, 2011
- The Regulation of Valvular and Vascular Sclerosis by Osteogenic MorphogensCirculation Research, 2011
- Crimping May Affect the Durability of Transcatheter Valves: An Experimental AnalysisThe Annals of Thoracic Surgery, 2011
- Quercetin-crosslinked porcine heart valve matrix: Mechanical properties, stability, anticalcification and cytocompatibilityActa Biomaterialia, 2010
- Detection of pathogenic Staphylococcus aureus bacteria by gold based immunosensorsMicrochimica Acta, 2008
- Hydrogel biomaterials: A smart future?Biomaterials, 2007
- In vitro re-endothelialization of detergent decellularized heart valves under simulated physiological dynamic conditionsBiomaterials, 2006
- Relevance of immunologic reactions for tissue failure of bioprosthetic heart valvesThe Annals of Thoracic Surgery, 1995