Vibratory insertion of press-fit acetabular components requires less force than a single blow technique
Open Access
- 5 June 2024
- journal article
- research article
- Published by British Editorial Society of Bone & Joint Surgery in Bone & Joint Research
- Vol. 13 (6) , 272-278
Abstract
Aims: Periprosthetic fracture and implant loosening are two of the major reasons for revision surgery of cementless implants. Optimal implant fixation with minimal bone damage is challenging in this procedure. This pilot study investigates whether vibratory implant insertion is gentler compared to consecutive single blows for acetabular component implantation in a surrogate polyurethane (PU) model.Methods: Acetabular components (cups) were implanted into 1 mm nominal under-sized cavities in PU foams (15 and 30 per cubic foot (PCF)) using a vibratory implant insertion device and an automated impaction device for single blows. The impaction force, remaining polar gap, and lever-out moment were measured and compared between the impaction methods.Results: Impaction force was reduced by 89% and 53% for vibratory insertion in 15 and 30 PCF foams, respectively. Both methods positioned the component with polar gaps under 2 mm in 15 PCF foam. However, in 30 PCF foam, the vibratory insertion resulted in a clinically undesirable polar gap of over 2 mm. A higher lever-out moment was achieved with the consecutive single blow insertion by 42% in 15 PCF and 2.7 times higher in 30 PCF foam.Conclusion: Vibratory implant insertion may lower periprosthetic fracture risk by reducing impaction forces, particularly in low-quality bone. Achieving implant seating using vibratory insertion requires adjustment of the nominal press-fit, especially in denser bone. Further preclinical testing on real bone tissue is necessary to assess whether its viscoelasticity in combination with an adjusted press-fit can compensate for the reduced primary stability after vibratory insertion observed in this study.Cite this article: Bone Joint Res 2024;13(6):272–278.Keywords
This publication has 37 references indexed in Scilit:
- Experimental investigations of the insertion and deformation behavior of press-fit and threaded acetabular cups for total hip replacementJournal of Orthopaedic Science, 2008
- The operation of the century: total hip replacementThe Lancet, 2007
- Artificial composite bone as a model of human trabecular bone: The implant–bone interfaceJournal of Biomechanics, 2007
- The number of screws, bone quality, and friction coefficient affect acetabular cup stabilityMedical Engineering & Physics, 2006
- Intraoperative Fractures of the Acetabulum During Primary Total Hip ArthroplastyJournal of Bone and Joint Surgery, 2006
- Deformation of press-fitted metallic resurfacing cups. Part 1: Experimental simulationProceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 2006
- Management of periprosthetic fractures: The hipThe Journal of Arthroplasty, 2002
- Press-fit stability of uncemented hemispheric acetabular components: a comparison of three porous coating systemsInternational Orthopaedics, 2002
- Effect of cementless acetabular cup geometry on strain distribution and press-fit stabilityThe Journal of Arthroplasty, 1997
- Periprosthetic Femoral FracturesJournal of the American Academy of Orthopaedic Surgeons, 1994