Creatinine Assay Attainment of Analytical Performance Goals Following Implementation of IDMS Standardization
Open Access
- 1 January 2017
- journal article
- research article
- Published by SAGE Publications in Canadian Journal of Kidney Health and Disease
- Vol. 4
Abstract
The international initiative to standardize creatinine (Cr) assays by tracing reference materials to Isotope Dilution Mass Spectrometry (IDMS) assigned values was implemented to reduce interlaboratory variability and improve assay accuracy. The aims of this study were to examine whether IDMS standardization has improved Cr assay accuracy (bias), interlaboratory variability (precision), total error (TE), and attainment of recommended analytical performance goals. External Quality Assessment (EQA) data (n = 66 challenge vials) from Ontario, Canada, were analyzed. The bias, precision, TE, and the number of EQA challenge vials meeting performance goals were determined by assay manufacturer before (n = 32) and after (n = 34) IDMS implementation. The challenge vials with the worst bias and precision were spiked with known common interfering substances (glucose and bilirubin). IDMS standardization improved assay bias (10.4%-1.6%, P < .001), but precision remained unchanged (5.0%-4.7%, P = .5) with performance goals not consistently being met. Precision and TE goals based on biologic variation were attained by only 29% to 69% and 32% to 62% of challenge vials. While IDMS standardization has improved Cr assay accuracy and thus reduced TE, significant interlaboratory variability remains. Contemporary Cr assays do not currently meet the standards required to allow for accurate and consistent estimated glomerular filtration rate assessment and chronic kidney disease diagnosis across laboratories. Further improvements in Cr assay performance are needed.Keywords
This publication has 19 references indexed in Scilit:
- Enzymatic but not compensated Jaffe methods reach the desirable specifications of NKDEP at normal levels of creatinine. Results of the French multicentric evaluationClinica Chimica Acta, 2013
- Recent Trends in Performance and Current State of Creatinine AssaysArchives of Pathology & Laboratory Medicine, 2013
- Can current analytical quality performance of UK clinical laboratories support evidence-based guidelines for diabetes and ischaemic heart disease? – A pilot study and a proposalClinical Chemistry and Laboratory Medicine (CCLM), 2013
- The Uncertainty of the eGFRIndian Journal of Clinical Biochemistry, 2012
- Screening for Chronic Kidney Disease: U.S. Preventive Services Task Force Recommendation StatementAnnals of Internal Medicine, 2012
- Specificity Characteristics of 7 Commercial Creatinine Measurement Procedures by Enzymatic and Jaffe Method PrinciplesClinical Chemistry, 2012
- A multicentric evaluation of IDMS-traceable creatinine enzymatic assaysClinica Chimica Acta, 2011
- Regional Implementation of Creatinine Measurement StandardizationJournal of the American Society of Nephrology, 2008
- Recommendations for Improving Serum Creatinine Measurement: A Report from the Laboratory Working Group of the National Kidney Disease Education ProgramClinical Chemistry, 2006
- Measurement of renal function in chronic renal diseaseKidney International, 1990