Quality control of radiotherapy treatment plans with electrons
Open Access
- 28 January 2019
- journal article
- Published by Sociedade Brasileira de Proteção Radiológica - SBPR in Brazilian Journal of Radiation Sciences
- Vol. 7 (1)
Abstract
In this work the quality of the treatment planning generated by EclipseÒ commercial treatment planning system for electron beams of energy 9 and 20 MeV was verified and their accuracy in the calculation of dose distributions for several clinical situations was evaluated. Tests cases had been developed according to the Brazilian reality, having as reference the radiotherapy treatments carried out in the National Institute of the Cancer.The system developed for checking the quality of treatment planning systems with electrons was efficient in evaluating the Eclipse planning system by identifying the failures of their algorithms, especially in planning the isodose. The verification system has been validated against the Monte Carlo method and the experimental data with an ionization chamber and showed the shortcomings of generalized pencil beam and eMC algorithms.Keywords
This publication has 11 references indexed in Scilit:
- Full Monte Carlo and measurement-based overall performance assessment of improved clinical implementation of eMC algorithm with emphasis on lower energy rangePhysica Medica, 2016
- Experimental verification and clinical implementation of a commercial Monte Carlo electron beam dose calculation algorithmMedical Physics, 2008
- Report of the AAPM Task Group No. 105: Issues associated with clinical implementation of Monte Carlo‐based photon and electron external beam treatment planningMedical Physics, 2007
- Impact of inhomogeneity corrections on dose coverage in the treatment of lung cancer using stereotactic body radiation therapyMedical Physics, 2007
- Review of electron beam therapy physicsPhysics in Medicine & Biology, 2006
- First macro Monte Carlo based commercial dose calculation module for electron beam treatment planning—new issues for clinical considerationPhysics in Medicine & Biology, 2006
- Comprehensive evaluation of a commercial macro Monte Carlo electron dose calculation implementation using a standard verification data setMedical Physics, 2006
- A model to determine the initial phase space of a clinical electron beam from measured beam dataPhysics in Medicine & Biology, 2000
- MMC-a high-performance Monte Carlo code for electron beam treatment planningPhysics in Medicine & Biology, 1995
- A macro Monte Carlo method for electron beam dose calculationsPhysics in Medicine & Biology, 1992