Calibration of Rogowski Coils With an Integrator at High Currents
- 25 April 2011
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Instrumentation and Measurement
- Vol. 60 (7) , 2434-2438
Abstract
This paper describes the calibration of Rogowski coils with an integrator for current measurement in the current range of up to 30 kA, at mains frequency. A current loop of ten turns with a current of 3 kA was used for calibration. The magnetic field generated in this way differs from the field of one turn placed at the center and passed with a current of 30 kA. The influence of the unevenness of winding and the cross section of the Rogowski coil may cause a calibration error when using a current loop. The difference in the results between using a current loop and one turn, and the measurement uncertainties are given here. Calculations of the mutual inductances of typical Rogowski coils and their uncertainties are presented in the Appendix.Keywords
This publication has 7 references indexed in Scilit:
- Calibration of Rogowski Coils With an Integrator at High CurrentsIEEE Transactions on Instrumentation and Measurement, 2011
- Calibration of Rogowski Coils at Power Frequencies Using Digital SamplingIEEE Transactions on Instrumentation and Measurement, 2009
- Onsite Calibration of a Current Transformer Using a Rogowski CoilIEEE Transactions on Instrumentation and Measurement, 2008
- Mutual Inductance of a Precise Rogowski Coil in Dependence of the Position of Primary ConductorIEEE Transactions on Instrumentation and Measurement, 2008
- A device for calibrating Rogowski coils in time domainReview of Scientific Instruments, 2008
- Influence of busbar geometry on AC current measurement using Rogowski coil2008
- Machinable Rogowski coil, design, and calibrationIEEE Transactions on Instrumentation and Measurement, 1996