Impact of the z-flying focal spot on resolution and artifact behavior for a 64-slice spiral CT scanner
- 6 April 2006
- journal article
- research article
- Published by Springer Nature in European Radiology
- Vol. 16 (6) , 1206-1215
Abstract
The effect of the z-flying focal spot (zFFS) technology was evaluated by simulations and measurements with respect to resolution and artifact behavior for a 64-slice spiral cone-beam computed tomography (CT) scanner. The zFFS alternates between two z-positions of the X-ray focal spot, acquiring two slices per detector row, which results in double sampling in the z-direction. We implemented a modified reconstruction that is able to obtain images as they would be without zFFS. A delta phantom equipped with a thin gold disc was used to measure slice sensitivity profiles (SSP), and a high-contrast bar phantom was used to quantify the resolution in the x/z-plane with and without zFFS. The zFFS decreases the full width at half maximum (FWHM) of the SSPs by a factor of about 1.4. The double z-sampling allows the separation of 0.4 mm bars in the z-direction compared with 0.6 mm in the case without zFFS. The zFFS effectively reduces windmill artifacts in the reconstructed images while maintaining the transverse resolution, even at the largest available pitch value of 1.5.Keywords
This publication has 8 references indexed in Scilit:
- Image reconstruction and image quality evaluation for a 64‐slice CT scanner with ‐flying focal spotMedical Physics, 2005
- Presampling, algorithm factors, and noise: Considerations for CT in particular and for medical imaging in generalMedical Physics, 2005
- New x‐ray tube performance in computed tomography by introducing the rotating envelope tube technologyMedical Physics, 2004
- The cause of the artifact in 4‐slice helical computed tomographyMedical Physics, 2004
- Windmill artifact in multislice helical CT2003
- Analytical models for multi-slice helical CT performance parametersMedical Physics, 2003
- Advanced single‐slice rebinning in cone‐beam spiral CTMedical Physics, 2000
- Algorithm for image reconstruction in multi-slice helical CTMedical Physics, 1998