Mammography radiation dose: initial results from Serbia based on mean glandular dose assessment for phantoms and patients
- 16 February 2010
- journal article
- research article
- Published by Oxford University Press (OUP) in Radiation Protection Dosimetry
- Vol. 140 (1) , 75-80
Abstract
The primary objective of this study is to investigate the actual mammography dose levels in Serbia and establish a baseline dose database for the future screening programme. The mean glandular dose (MGD) was assessed for standard size breast substituted by 45 mm polymethyl methacrylate (PMMA) phantom and imaged under typical clinical conditions at 30 screen-film mammography facilities. Average MGD was (1.9 ± 1.0) mGy, with a range of 0.12–5.2 mGy. These results were used for the calculation of the initial diagnostic reference levels in mammography in Serbia, which was set to 2.1 mGy for a standard breast. At some facilities, high doses were observed and the assessed MGD was supplemented by a patient dose survey. MGD was measured for 53 women having a diagnostic mammography in cranio-caudal (CC) and mediolateral oblique (MLO) projections at the largest mammography facilities in Serbia. The parameters recorded were: age, compressed breast thickness, tube potential (in kV), tube loading (in mAs) and the MLO projection angle. The average MGD per image was 2.8 mGy for the CC projection and 4.3 mGy for the MLO projection. Dose to breast was compared with dose for PMMA phantoms of various sizes, assessed at the same facilities. The results have indicated that phantom dose values can assist in setting optimisation activities in mammography. Both phantom and patient dose values indicated unnecessary high doses in facilities with an extremely high mammography workload. For these facilities, urgent dose-reduction measures and follow-up actions were recommended.Keywords
This publication has 15 references indexed in Scilit:
- Investigation of breast dose in five screening mammography centres in GreeceJournal of Radiological Protection, 2008
- Survey of mammography practice in Croatia: equipment performance, image quality and doseRadiation Protection Dosimetry, 2008
- Evaluation of the population dose to the UK population from the National Health Service breast screening programmeRadiation Protection Dosimetry, 2008
- Results of a European dose survey for mammographyRadiation Protection Dosimetry, 2008
- Reference levels for image quality in mammographyRadiation Protection Dosimetry, 2008
- Investigation of mean glandular dose versus compressed breast thickness relationship for mammographyRadiation Protection Dosimetry, 2008
- Global quality control perspective for the physical and technical aspects of screen-film mammography—image quality and radiation doseRadiation Protection Dosimetry, 2006
- Towards a proposition of a diagnostic (dose) reference level for mammographic acquisitions in breast screening measurements in BelgiumRadiation Protection Dosimetry, 2005
- A study of mean glandular dose during diagnostic mammography in Malaysia and some of the factors affecting itBritish Journal of Radiology, 2003
- Breast compression and radiation dose in two different mammographic oblique projections: 45 and 60°European Journal of Radiology, 2001