Comparison of Goldmann applanation tonometry and Pascal dynamic contour tonometry in relation to central corneal thickness and corneal curvature
Open Access
- 1 December 2016
- journal article
- research article
- Published by Taylor & Francis in Clinical Ophthalmology
- Vol. ume 10, 2477-2484
Abstract
Purpose: The aim of the study was to assess the influence of central corneal thickness (CCT) and corneal curvature in tonometry measurements taken by Goldmann applanation tonometry (GAT) and Pascal dynamic contour tonometry (DCT). Methods: This was a prospective study of 185 eyes from 97 subjects, attending outpatient ophthalmology appointments, who underwent intraocular pressure measurements by GAT and Pascal DCT. CCT and corneal curvature were obtained using ultrasound pachymetry and Orbscan topography, respectively. All measurements were carried out among males and females during the period 2009–2012. Apart from the usual descriptive and exploratory data analysis, one-way analysis of variance and agreement analysis were performed, linear as well as intraclass correlation coefficients were estimated, and multiple scatter and Bland–Altman plots were produced. Results: Mean IOP measurements obtained were 17.21±4.10 mmHg by DCT and 13.23±4.07 mmHg by GAT. Mean difference between the GAT and DCT measurements was 3.88±2.8 mmHg. Mean CCT and corneal curvature were 522.78±52 µm and 43.83±2.9823 D, respectively. Conclusion: Intraocular pressure measured by GAT was consistently lower when compared with DCT, and this difference was greatest with thinner CCT. Flat corneas seem to influence GAT measurements compared to DCT.Keywords
This publication has 30 references indexed in Scilit:
- Comparison of Goldmann and Pascal tonometry in relation to corneal hysteresis and central corneal thickness in nonglaucomatous eyesClinical Ophthalmology, 2011
- Comparison of dynamic contour tonometry with Goldmann applanation tonometry in glaucoma practiceActa Ophthalmologica, 2009
- Comparison Between Dynamic Contour Tonometry and Goldmann Applanation TonometryKorean Journal of Ophthalmology, 2009
- Influence of age, central corneal thickness, and quality score on dynamic contour tonometryEye, 2008
- Corneal Parameters and Difference Between Goldmann Applanation Tonometry and Dynamic Contour Tonometry in Normal EyesJournal of Glaucoma, 2008
- Comparison of Goldmann applanation tonometry and dynamic contour tonometry in healthy and glaucomatous eyesEye, 2008
- The relative effects of corneal thickness and age on Goldmann applanation tonometry and dynamic contour tonometryBritish Journal of Ophthalmology, 2005
- Comparison of intraocular pressure measured by Pascal dynamic contour tonometry and Goldmann applanation tonometryEye, 2005
- Central corneal thickness, radius of the corneal curvature and intraocular pressure in normal subjects using non‐contact techniques: Reykjavik Eye StudyActa Ophthalmologica Scandinavica, 2002
- Über ApplanationstonometrieOphthalmologica, 1957