Automated control of end-tidal inhalation anaesthetic concentration using the GE Aisys Carestation™ †
Authors:
- P. BarclayPBP. Barclay
- 1 April 2013
- journal article
- research article
- Published by Elsevier in British Journal of Anaesthesia
- Vol. 110 (4) , 561-566
Abstract
Automated control of end-tidal inhalation anaesthetic concentration is now possible. The EtControl™ module of an Aisys Carestation Anaesthetic machine digitally adjusts fresh gas flow and plenum vaporizer output to achieve a target end-tidal concentration. We evaluated EtControl in clinical practice by measuring volatile agent consumption and the need for user input. We compared these values with contemporaneous controls using manual control of fresh gas flow rates. A total of 321 patients were anaesthetized with EtControl and 168 with manual control of fresh gas flow. The mean [95% confidence interval (CI)] sevoflurane usage for cases of 20-40 min duration was 14 (13-16) ml h(-1) with EtControl and 30 (26-35) ml h(-1) with manual control. For cases of the same duration, the mean (95% CI) desflurane consumption was 27 (21-33) ml h(-1) with EtControl and 45 (29-62) ml h(-1) with manual control. The average number of keypresses per case was 6.5 with EtControl and 13.6 during manual control of fresh gas flow. Automatic implementation of low-flow anaesthesia using EtControl allows the user to set and maintain a desired end-tidal volatile concentration while using less volatile agent.Keywords
This publication has 14 references indexed in Scilit:
- Effect of Low‐flow Anesthesia Education on Knowledge, Attitude and Behavior of the Anesthesia TeamThe Kaohsiung Journal of Medical Sciences, 2010
- Assessing the clinical or pharmaco‐economical benefit of target controlled desflurane delivery in surgical patients using the Zeus ® anaesthesia machine*Anaesthesia, 2009
- Target‐controlled inhalation induction with sevoflurane in children: a prospective pilot studyPediatric Anesthesia, 2009
- Changing Patterns in Anesthetic Fresh Gas Flow Rates Over 5 Years in a Teaching HospitalAnesthesia & Analgesia, 2008
- Fresh gas flow is not the only determinant of volatile agent consumption: a multi-centre study of low-flow anaesthesia † †Conflict of interest: This study was funded by a grant from Daäger South Africa (Pty) Ltd. and coordinated by the Technology and Business Development Group of Medical Research Council of South Africa (MRC).British Journal of Anaesthesia, 2002
- Measuring the costs of inhaled anaestheticsBritish Journal of Anaesthesia, 2001
- Model-based automatic feedback control versus human control of end-tidal isoflurane concentration using low-flow anaesthesiaBritish Journal of Anaesthesia, 2000
- Individualized Feedback of Volatile Agent Use Reduces Fresh Gas Flow Rate, but Fails to Favorably Affect Agent ChoiceAnesthesiology, 1999
- The uptake of sevoflurane during anaesthesiaAnaesthesia, 1998
- Uptake of desflurane during anaesthesiaAnaesthesia, 1996