Comparison of 6 Oscillatory Positive Expiratory Pressure Devices During Active Expiratory Flow
Authors:
- 19 November 2019
- journal article
- research article
- Published by SAGE Publications in Respiratory Care
- Vol. 65 (4) , 492-499
Abstract
BACKGROUND: Air-flow oscillations generated by exhaling through oscillatory positive expiratory pressure (OPEP) devices favor airway clearance. Variations in mechanical properties between different devices may influence therapeutic efficacy. The objective of this study was to assess mechanical properties in vitro and to compare the performance of 6 OPEP devices at different resistance levels under active expiratory flow patterns. METHODS: 4 gravity-dependent OPEP devices (ie, Flutter, Gelomuc, Pari O-PEP, Shaker Medic Plus) and 2 gravity-independent OPEP devices (ie, Acapella Choice and Aerobika) were each tested at low, medium, and high resistance settings. All devices were independently connected to a pulmonary waveform generator that reproduced active exhalation flows. Expiratory flow-volume curves were retrieved from 4 subjects with different stages of obstruction severity and were scaled according to either peak expiratory flow (4, 6, and 8 L/s) or volumes (2, 3 and 4 L), thus amounting to 24 active exhalations. Resulting waveforms were divided into 4 parts and the 2 middle parts were used to extract the following mechanical data: positive expiratory pressure (PEP), maximum expiratory pressure (P peak ), oscillation frequency, and flow oscillation amplitude. The percentage of tests achieving oscillation frequencies ≥ 12 Hz and PEP ≥ 10 cm H 2 O was calculated for each device. RESULTS: Mechanistic effects of the Acapella, Aerobika, and Shaker devices were not comparable. The Flutter, Gelomuc, and Pari devices behaved similarly and achieved more tests with optimum oscillation frequency and PEP values than the other devices. These 3 devices also produced the highest oscillation amplitudes at the low-resistance level, whereas the Aerobika elicited higher and consistent oscillation amplitudes at medium and high resistance settings. CONCLUSIONS: Operational parameters differed between and within devices, yet the Flutter, Gelomuc, and Pari devices were similar in many aspects. Therapeutic efficacy may depend on the selected OPEP device and set resistance.Keywords
This publication has 26 references indexed in Scilit:
- Evaluation of Functional Characteristics of 4 Oscillatory Positive Pressure Devices in a Simulated Cystic Fibrosis ModelRespiratory Care, 2017
- Oscillatory Positive Expiratory Pressure in Chronic Obstructive Pulmonary DiseaseCOPD: Journal of Chronic Obstructive Pulmonary Disease, 2015
- Positive expiratory pressure – Common clinical applications and physiological effectsRespiratory Medicine, 2015
- Laboratory Evaluation of Four Different Devices for Secretion Mobilization: Acapella Choice, Green and Blue Versus Water BottleRespiratory Care, 2013
- Structure and Function of the Mucus Clearance System of the LungCold Spring Harbor Perspectives in Medicine, 2013
- Reference Values of Impulse Oscillometric Lung Function Indices in Adults of Advanced AgePLOS ONE, 2013
- Relationship Between Spontaneous Expiratory Flow-Volume Curve Pattern and Air-Flow Obstruction in Elderly COPD PatientsRespiratory Care, 2013
- Mechanical Behaviors of Flutter VRP1, Shaker, and Acapella DevicesRespiratory Care, 2013
- Transient reference values for impulse oscillometry for children aged 3–18 yearsPediatric Pulmonology, 2008
- Mechanical evaluation of a respiratory deviceMedical Engineering & Physics, 2004