Oxygen electrode dynamics: Three‐layer model—chemical reaction in the liquid film
Authors:
- 1 May 1983
- journal article
- research article
- Published by Wiley in Biotechnology and Bioengineering
- Vol. 25 (5) , 1401-1418
Abstract
A general model of the transient behaviour of an oxygen probe is derived on more general assumptions than those previously considered. The present state of the art is shorly discussed in its relation to the newly derived equations: (1) the reducibility of the three‐layer diffusional model to two‐layer ones is discussed, and the two‐layer model considering the membrane and the liquid film is verified experimentally, (2) experimental and theoretical treatments of transient probe characteristics determination are examined, and (3) the influence of chemical reactions of the first order as well as of zero order (microorgaganism respiration) in measured media on transientd probe characteristics is quantified.Keywords
This publication has 17 references indexed in Scilit:
- Dynamic measurement of the overall volumetric mass‐transfer coefficient in air‐sparged systemsBiotechnology and Bioengineering, 1979
- Oxygen Probe Dynamics in Flowing FluidsIndustrial & Engineering Chemistry Fundamentals, 1979
- Error in oxygen measurements in open systems owing to oxygen consumption in unstirred layerBiotechnology and Bioengineering, 1978
- Dynamic measurement of the volumetric mass transfer coefficient in agitated vessels: Effect of the start‐up period on the response of an oxygen electrodeBiotechnology and Bioengineering, 1977
- Oxygen transfer coefficients by dynamic model moment analysisBiotechnology and Bioengineering, 1977
- Oxygen electrode response lag induced by liquid film resistance against oxygen transferBiotechnology and Bioengineering, 1976
- Determination of aeration capacity of mechanically. Agitated vessels by fast response oxygen probeBiotechnology and Bioengineering, 1972
- Oxygen mass transfer and oxygen respiration rate measurements utilizing fast response oxygen electrodesBiotechnology and Bioengineering, 1971
- Polarographic oxygen analyzer response: The effect of instrument lag in the non-steady state reaeration testWater Research, 1970
- On the use of fast‐response dissolved oxygen probes for oxygen transfer studiesBiotechnology and Bioengineering, 1970