Intra-DIC (distal intracranial catheter) deployment of the Pipeline embolization device: a novel rescue strategy for failed device expansion
- 22 June 2015
- journal article
- case report
- Published by BMJ in Journal of NeuroInterventional Surgery
- Vol. 8 (8) , 840-846
Abstract
Background The Pipeline embolization device (PED) is a braided flow diverter that requires a combination of meticulous maneuvers to assure proper device opening and expansion. Mechanical, anatomical, or technical challenges can result in a partially deployed PED with failed expansion. Objective To present a new alternative method of PED deployment using the Navien distal intracranial catheter (DIC) as a salvage maneuver for cases where PED opening fails with standard techniques. Methods We retrospectively reviewed a prospective, single-center aneurysm database to identify all patients who underwent endovascular treatment of intracranial aneurysms using the PED with the Navien distal intracranial catheter access platform. Cases requiring PED deployment within the Navien catheter were reviewed. Data was collected for patient demographics, aneurysm characteristics, and technical details of the interventional procedure. Results Eleven PED neurointerventions requiring intra-Navien PED deployment to fully open the PED were identified. Mean patient age was 55.5±9.9 years (range 37–76 years). Mean aneurysm size was 12.5 mm±4.9 mm (range 2–42 mm). All aneurysms were located in the anterior circulation (anterior cerebral artery, n=1; supraclinoid, n=1; ophthalmic/paraophthalmic, n=6; cavernous, n=3; petrocervical, n=1). Mean fluoroscopy time was 67.1±20.5 min. The intra-Navien technique was used to open the proximal PED (n=7) and the mid-portion (n=4). Post-processing of the PED with a balloon was used in six cases. Conclusions When a partially deployed PED remains constrained despite exhaustion of standard maneuvers to facilitate opening, the technique of intra-Navien PED deployment is a valuable rescue strategy. This new alternative method of PED deployment can be used to open a stretched device with successful completion of the PED implantation.Keywords
This publication has 11 references indexed in Scilit:
- Pipeline Embolization Device for the Treatment of Intra- and Extracranial Fusiform and Dissecting AneurysmsNeurosurgery, 2014
- Classification of cavernous internal carotid artery tortuosity: a predictor of procedural complexity in Pipeline embolizationJournal of NeuroInterventional Surgery, 2014
- Retrograde trans-anterior communicating artery rescue of unopened Pipeline Embolization Device with balloon dilation: complication managementJournal of NeuroInterventional Surgery, 2014
- A Paradigm-Shifting Technology for the Treatment of Cerebral Aneurysms: The Pipeline Embolization DeviceWorld Neurosurgery, 2013
- Endovascular Treatment of Middle Cerebral Artery Aneurysms with Flow Modification with the Use of the Pipeline Embolization DeviceAmerican Journal of Neuroradiology, 2013
- Utilization of the Navien distal intracranial catheter in 78 cases of anterior circulation aneurysm treatment with the Pipeline embolization deviceJournal of NeuroInterventional Surgery, 2013
- Treatment of Intracranial Aneurysms Using the Pipeline Flow-Diverter Embolization Device: A Single-Center Experience with Long-Term Follow-Up ResultsAmerican Journal of Neuroradiology, 2012
- Safety of the Pipeline Embolization Device in Treatment of Posterior Circulation AneurysmsAmerican Journal of Neuroradiology, 2012
- Immediate procedural outcomes in 35 consecutive pipeline embolization cases: a single-center, single-user experienceJournal of NeuroInterventional Surgery, 2012
- In situ removal of the pipeline embolization device: the ‘corking’ and ‘pseudo-corking’ techniquesJournal of NeuroInterventional Surgery, 2012